41
remote interface operation.
Front Panel Operation
1.
Press ACV button and then CONFIG buttons as shown in
Figure 3-2
.
2.
Use
◁
and
▷
buttons to
locate “BAND WIDTH” submenu, and then press ENTER
to select it. (There are three options: 3Hz, 20Hz and 200Hz.)
Figure 3-2
Remote Interface Operation
From your PC terminal, use the following command to specify the filter type:
DETector:BANDwidth {3|20|200|MIN|MAX}
3.1.2.2 Digital Filter
Definition:
DL-1060 uses an averaging digital filter to yield a reading for display from a specified
number of measurement readings in the past. The past measurement readings are
stored in a stack of memory. The number may be in the range of 2 to 100. You
may select one of the two modes of digital filter operations: Moving Average and
Repeating Average.
The moving average filter puts the specified number of reading conversions in a
first-in, first-out order. The very first measurement reading simply fills up the stack.
To yield a reading for display, the filter produces an average of the stacked
measurement readings every time when a new measurement reading is available and
replaces the oldest reading in the stack. In the repeating average mode, the
multimeter waits for the measurement reading stack to fill up and then take an
average to produce a reading for display. It then flushes the stack and starts over
with an empty stack. Consequently, the repeating digital filter yields one reading for
display every specified number of measurement readings.
※
Note: Digital filter is not available for diode, continuity, frequency and period
measurements.