12
AUTO
Indicates Automatic range selection
10A
2nd display shows the Current range
3. Connect the test lead
and measure
Connect the test lead between the A
and COM port or LO to COM port,
depending on the current. For current ≤
2A use the LO port; For current up to
10A use the A port. The display
updates the reading.
A
COM
LO
MAX
2A
2-10. Select Current range
Auto range
To turn the automatic range selection On/Off, press
the AUTO key.
Manual range
Press the Up or the Down key to select the range.
AUTO indicator turns Off automatically. If the
appropriate range is unknown, select the highest
range.
Selection list
Range
Resolution / Full scale @ slow rate
Resolution
Full scale
10mA
0.1µA
12.0000mA
100mA
1µA
120.000mA
1A
100µA
1.2000A
10A
100µA
10.0000A
Note
*10A range is not available for AC+DC Current.
For more detailed range, see the specifications at page50.
2-11. 2W/4W Resistance Measurement
Measurement type
2-wire
Uses the standard V-COM ports.
Recommended for measuring resistances larger than 1kΩ.
4-wire
Compensates the test lead effect using the 4W compensation
ports, in addition to the standard V-COM ports.
Recommended for measuring sensitive resistances smaller
than 1kΩ.
1. Activate resistance
measurement
For 2-wire resistance measurement, press the
2W/4W key once.
For 4-wire resistance measurement, press the
2W/4W key twice.
2. 2W resistance mode
display appears
2W(4W) + Ω
Indicates 2W(4W) Resistance mode
AUTO
Indicates Automatic range selection
10M
2nd display shows the Resistance range
3. Connect the test lead
and measure
Connect the test lead. For 2-wire res
istance, use the Ω (V) and the COM port.
For 4-
wire resistance, use the Ω (V) and the COM port, plus the 4W sense, and
LO port for sensing. The display updates the reading.