52
9-8. 4W Resistance
Note
Max. Input: 500V DC or 500V rms AC
Rate
Range
Full Scale
Accuracy rdigits
S
100.000
Ω
120.000
Ω
0.05% + 8
1.00000k
Ω
1.20000k
Ω
0.05% + 5
10.0000k
Ω
12.0000k
Ω
0.05% + 5
100.000k
Ω
120.000k
Ω
0.05% + 5
1.00000M
Ω
1.20000M
Ω
0.05% + 5
10.0000M
Ω
12.0000M
Ω
0.3% + 5
100.000M
Ω
120.000M
Ω
3.0% + 8
M
100.00
Ω
120.00
Ω
0.05% + 5
1.0000k
Ω
1.2000k
Ω
0.05% + 3
10.000k
Ω
12.000k
Ω
0.05% + 3
100.00k
Ω
120.00k
Ω
0.05% + 3
1.0000M
Ω
1.2000M
Ω
0.05% + 3
10.000M
Ω
12.000M
Ω
1.5% + 3
100.00M
Ω
120.00M
Ω
5.0% + 5
F
100.0
Ω
120.0
Ω
0.05% + 2
1.000k
Ω
1.200k
Ω
0.05% + 2
10.00k
Ω
12.00k
Ω
0.05% + 2
100.0k
Ω
120.0k
Ω
0.05% + 2
1.000M
Ω
1.200M
Ω
0.05% + 2
10.00M
Ω
12.00M
Ω
1.5% + 2
100.0M
Ω
120.0M
Ω
5.0% + 2
9-9. Diode/Continuity
Note
Max. Input: 500V DC or 500V rms AC
Item
Range
Diode
Approx. 2V, 0.5mA
Continuity
1 ~ 1000
Ω
9-10. Frequency
Note
Max. Input: 750V rms or 1000V peak
Frequency
Sensitivity
Accuracy (rdigits)
10Hz ~ 100kHz
0.1V
0.05% + 15
100kHz ~ 600kHz
1V
0.05% + 3
600kHz ~ 800kHz
2.5V
0.05% + 3
9-11. Temperature
Note
Sensor errors excluded from Temperature
specifications
Type
Measurement Range
Thermo Couple
K
0 ~ +300°C
T
0 ~ +300°C
J
0 ~ +300°C
Resolution
0.01°C (0 ~ 300°C)
9-12. Accessories
Accessories CD-ROM
Instruction
CAL KEY
GDM-01
Test leads
GTL-107 or GTL-207