21
dB result appears
dB
Indicates dB measurement
113.729mV
Indicates the present Voltage reading
dBmref
Press the 2
ND
key to see the dBmref value.
Deactivate dB
measurement
To cancel the dBm measurement, press the Shift
key followed by the Hz/P key, or simply activate
another measurement.
3-8. Max/Min Measurement
Applicable to
( )
+
DCV
ACI
DCI
2/4W
ACV
Hz/P
TEMP
( )
+
Background
Maximum and Minimum measurement stores the highest (maximum) or lowest
(minimum) reading and shows it on the 2nd display.
1. Activate Max/Min For Max measurement, press the MX/MN key once.
For Min measurement, press the MX/MN key twice.
2. Max (Min) result
appears
MIN (MAX)
Indicates Min (Max) measurement
0.11516
2nd display shows the Min (Max) measurement result
Deactivate
Max/Min
measurement
To cancel the Max/Min measurement, press the MX/MN
key for 2 seconds, or simply activate another
measurement.
3-9. Relative Value Measurement
Applicable to
( )
+
DCV
ACI
DCI
2/4W
ACV
Hz/P
TEMP
( )
+
Background
Relative measurement stores a value, typically the data at the moment, as the
reference. The following measurement is shown as the delta between the reference.
1. Activate
Relative
measurement
Press the REL key. The measurement reading at the
moment becomes the reference value.
2. Relative
measurement
display appears
REL
Indicates Relative value measurement
2nd display Shows the reference value
1st display
Shows the delta between the current measurement data and the
reference value
Manually set the
reference value
1. To set the reference value manually, press the
Shift key followed by the REL key. The setting
appears.