148
Note
For ramp waveforms, the Apply command and AM/FM
modulation modes ignore the current symmetry settings.
Syntax
SOURce[1|2]:RAMP:SYMMetry {< percent>
|MINimum|MAXimum}
Example
SOUR1:RAMP:SYMM MAX
Sets the symmetry to the 100%.
Query Syntax
SOURce[1|2]:RAMP:SYMMetry? {MINimum|MAXimum}
Return
Parameter
<NR3>
Returns the symmetry as a percentage.
Example
SOUR1:RAMP:SYMMetry?
+1.0000E+02
The symmetry is set as 100%.
8-7-7. OUTPut[1|2]
Set
Query
Description
Enables/Disables or queries the front panel output from the
selected channel. The default is set to off.
Note
If the output is overloaded by an external voltage, the output
will turn off and an error message will be displayed. The
overload must first be removed before the output can be
turned on again with output command.
Using the Apply command automatically sets the front panel
output to on.
Syntax
OUTPut[1|2] {OFF|ON}
Example
OUTP1 ON
Turns the channel 1 output on.
Query Syntax
OUTPut[1|2]?
Return
Parameter
1
0
ON
OFF
Example
OUTP1?
1
The channel 1 output is currently on.