96
Power Source
OPP time out
OP
P time out will occur if the power supply’s
OPP fails to trigger. This is determined when
the measured voltage is less than Trig V and
the measured current is greater than End W.
Power Source
Config Error
Config Error indicates that the power supply
voltage is less than the Trig V voltage setting
after the test has started. This can indicate that
the power supply output is not on or that the
power supply output or Trig V is incorrectly
configured.
Note
In addition to the OPP settings as described
above, the VON voltage settings must also be set
according to the output characteristics of the DUT.