99
Setting
Sets the values corresponding to the
defined discharging mode (CC mode
in A, CR mode in mS and CP mode
in W).
SlewRate
=
Sets the test rising slew rate in
mA/us (not adjustable for CP mode).
SlewRate
?
Sets the test falling slew rate in
mA/us (not adjustable for CP mode).
Stop Volt
Sets the voltage at which the test
should be interrupted. The value
must be lower than the battery start
voltage.
Stop Time
Sets the time after which the test
should be interrupted (max value is
999h:59m:59s).
Stop AH
Sets the discharged energy rate at
which the test should be interrupted
(Max value is 9999.99Ah).
Datalog timer Sets the time interval for data
capture.
Panel operation
1. Press
FUNC
>
Next Manu[F5].
>