35
Operation
1.
The short function can be turned on and off by
pressing the
Short
key.
The
Short
key will turn red when the short
function is active.
The Short icon will appear when the short
function is active.
Range:
Toggle, Hold
Display
Mode
CP
IRange
H 35A
VRange
L 15V
Function
Dynamic
Configure
1.500
V
0.000
A
1.50
w
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
SHORT
Fine
31/Oct/2013
mS
W
W
SHORT on
Note
If the load is already off, pressing the Short key
will turn the load on (shorted) at the same time.
Pressing the Short key again will also turn the
load off again as well.
If the load is already on and the Short key is
pressed, then when the Short key is pressed
again the load will remain on (the electronic load
will return to its previous load condition).
2-1-9. Safety Short
Description
When activated, the safety short function only
allows the short key to be used when the load is
already on.
Operation
1. Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Short Safety
.
When set to OFF, the load can be shorted at
anytime.
When set to ON, the load can only be shorted
when the load is already on.
Short(Safety)
:
OFF,ON
Note
The Short Safety setting will be grayed out if Short
Function is set to OFF.