7-5-3. CP Mode
CP Mode
When the unit is set to CP mode it will operate
as a constant power load when connected to a
constant voltage source. This means the unit will
maintain a set power level, up to the rated
current or voltage level, regardless of the input
voltage. When input voltage changes, the unit
responds by changing the current load to
maintain the set power level accordingly (P=IxV).
This is illustrated below.
CP Mode
Load Current
Load Input
Voltage
CP Value
CP+CV Mode
When CP+CV mode is enabled, the unit will act
as a constant power load after the input voltage
is greater than the user-defined CV level. At the
CV level, the unit works as a constant voltage
load. This mode effectively creates a voltage
ceiling before the unit operates in CP mode. The
diagram below illustrates this.
CP+CV Mode
Load Current
Load Input
Voltage
CP Value
CV level
Note that when the source voltage is less than
the CV level, no current will flow due to a very
high impedance.