32
2-1-5. CP Mode
Description
In Constant Power Mode, the unit will maintain a
constant power by varying the current. For more
details on CP mode, see the appendix on page
135.
Warning
If you change the mode or the range when the
load is already on, the load will be turned off
automatically.
Operation
1.
Make sure the load is off.
2.
Press
Main
.
3.
Select CP mode with the
Mode[F1]
soft-key.
4.
Select the current range with the
I Range[F2]
soft-key.
Range:
High, Middle, Low
5.
Select the voltage range with the
V Range[F3]
soft-key.
Range:
High, Low
6.
Set the power level parameters using the scroll
wheel and number pad.
For Static mode, set
CP A Value
and/or
CP B
Value
.
For Dynamic mode, set
Level1
and
Level2
.
The maximum and minimum power levels
depend on the selected current range.
For static mode, the parameter that is set last
becomes the “active” setting. This will be
shown in the Operation Status Panel.
7.
To add CV mode to CP mode (CP+CV), see
page 33.
8.
Set the remaining basic configuration settings
such as the slew rate, and timer settings. See
page 37 for details.