2-3-3-2. Cut Off Time
Description
The Cut Off Time function will turn the load off
after a set-amount of time. After the load has
been turned off, a popup screen will display the
voltage level when the load was turned off.
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Cut Off Time
.
Range:
OFF, 1 second
–
999 hours:59minutes:59 seconds
Display
Mode
CP
I Range
H 35A
V Range
L 15V
Function
Dynamic
Configure
0.000
V
0.000
A
0.00
w
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
Fine
31/Oct/2013
mS
W
W
0:00:05
Cut off time
Time Up
Voltage : 5.1223V
Enter
Voltage at cut off time
2-3-4. Auto Load Configuration
Description
The LSG Series can be configured to
automatically load the last program, normal
sequence, fast sequence or load setting at
startup.
By default, this setting is disabled.
Operation
1.
Press
Shift
+
Utility
Help
>
Load[F2]
.
2.
Turn
Auto Load
On or Off.
When set to OFF, the Auto Load setting is
disabled.
3.
Select the
Auto Load On
configuration.
This will select whether the LSG Series will
automatically load the last program, normal
sequence, fast sequence or load settings.
Auto Load On:
Load, Prog, NSeq, FSeq