background image

CL-B101D40-TCRTD Manual (d0099)

Page 7

Operational Display

Sub-menu 
MODE

STEP A  Calibration

Mode

STEP B  Calibration

Mode

 

STEP E

Calibrate 

Analog 

Output 

Lo 

STEP F

Calibrate 

Analog 

Output 

Hi 

Will only appear if the analog
output option is installed

To Digital Span Selection

for Bargraph Display

See Page 5

To Thermocouple

or RTD Selection

Mode See Page 4

Operational Display

TWO POINT ANALOG OUTPUT

RANGE SETTING AND CALIBRATION

+

x

x

4.00

+

x

x

20.00

Determine if the Analog Output Selection Header is in the 4 to 20mA 

(0-20mA) position or the 0 to 10VDC position.  If necessary, the mod-

ule may have to be removed and the header position changed (see 

Component Layout below).

Note

: Always disconnect power from the meter before removing the 

analog output module to adjust the mA or Volts output selection head-

er and reinstalling it.  When power is reconnected, the meter’s software 

will automatically detect the presence or absence of the analog output 

module.

STEP A  Enter the Calibration Mode

 1) 

Press the 

P

 and   buttons at the same time. Display 

toggles between [cAL] and [oFF].

 2) 

Press the   or   button. Display changes from [oFF] to 

[on].

 3) 

Press the 

P

 button. Display toggles between [cAL] and 

[out] input calibration.

Note

: If at this point the display skips directly to toggle between 

Zero and the previous Zero setting, the software is detecting that the 

optional analog output hardware is NOT installed.  

STEP B  Enter the Two Point Analog [ouT] Output Range Setting 

and Calibration Mode

 

1)  Press the 

P

 button. Display toggles between [cLo] and an 

internal scale factor.

STEP E  Set or Calibrate [cLo] the Low Analog Value of the 

Analog Output Range

 

1)  Connect a multimeter to analog output pins 17 and 18 

(see Rear Panel Pinouts on page 10). Using the   and   

buttons, adjust the analog output to the desired low value as 

measured on the multimeter. cLo may be adjusted to any 

value from –0.3 mA to 18 mA (mA output selected) or from 

–0.6 V to 8 V (volt output selected).  However, the output of 

cLo must always be less than the value selected for chi.  If a 

reversed analog output is desired, the values selected to 

establish the Digital Span can be reversed (see top of page 

6).  For digital readings outside the Digital Span selected, the 

analog output will not go any lower than the calibrated value 

set for cLo.  However, the analog output will linearly rise 

above the value set for chi, up to the the maximum analog 

output capability (see chi below).

 

2)  Press the 

P

 button. Display toggles between [chi] and an 

internal scale factor.

STEP F  Set or Calibrate [chi] the High Analog Value of the 

Analog Output Range

 

1)  Using the   and   buttons, adjust the analog output 

to the desired high value as measured on the multimeter 

display. chi may be adjusted to any value from 18 mA to 

24 mA (mA output) or from 8 V to 10.3 V (volt output). 

However, the value must be higher than the value select-

ed for cLo.  For digital readings outside the Digital Span 

selected, the analog output will linearly rise above the 

value set for chi, up to the maximum analog output capa-

bility.

 

2)  Press the 

P

 button. The meter exits the calibration 

mode and returns to the operational display.

Note:

 The analog output range established by the values selected 

for cLo and chi will occur, automatically scaled, between the two 

digital values selected for AnHi and AnLo.  However, the analog 

output can linearly rise above the chi value set for digital readings 

outside the digital span selected.  See Digital Span Selection on 

page 6.

Two Point Analog Output Range Setting and Calibration

Two Point Analog Output Range Setting and Calibration

Summary of Contents for CL-B101D40-RTD

Page 1: ...or below Lo each setpoint Digital display blanking Decimal point setting Four level brightness control LEOPARD FAMILY Input Specs Depends on Input signal conditioner A D Converter 14 bit single slope...

Page 2: ...direction indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are sho...

Page 3: ...999 seconds Pump PUM Hysteresis hYSt Hysteresis hYSt Hysteresis hYSt Setpoint 3 SP3 Setpoint 4 SP4 Operational Display MAIN MENU BRIGHTNESS AND SETPOINT VIEW MODE Bargraph Center Point Display Selecti...

Page 4: ...E STEP A Calibration Mode STEP B Calibration Mode To Digital Span Selection for Bargraph Display Page 5 cAL out iP branch will only appear if the analog output option is installed otherwise cAL on goe...

Page 5: ...between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph Display Bargraph does not light up for Input Signals up to...

Page 6: ...igh and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Smal...

Page 7: ...nd 18 see Rear Panel Pinouts on page 10 Using the and buttons adjust the analog output to the desired low value as measured on the multimeter cLo may be adjusted to any value from 0 3 mA to 18 mA mA o...

Page 8: ...Press the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value...

Page 9: ...Bargraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color se...

Page 10: ...An analog output module is also shown as installed The CL B101D40 TCRTD uses plug in type screw terminal con nectors for all input and output connections The power supply connections pins 14 and 15 h...

Page 11: ...A 5A 5A 5A 5A 5A 5A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog O...

Page 12: ...the back panel back into the case Apply power to the meter 4 Enter the program mode and select the type of thermocouple J K R T the resolution 0 1 or 1 and the display units C or F See Page 4 of the d...

Page 13: ...10 72VDC Input Modules Partial List See www texmate com IT10 Thermocouple J K R T Selectable C F 1 0 1 IT11 RTD 100 Pt Selectable 3 4 wire C F 1 0 1 385 392 IT15 RTD 1000 Pt Selectable 3 4 wire C F 1...

Reviews: