When the rear Auto/Manual switch is in the Auto position, the meter can be calibrated  with an automatic  scale factor calculation, 

by applying  a low input signal (InLo), entering the desired low display reading for that signal (ZERO),  then applying  a high input 

signal (InHi), and then entering the desired high display reading (SPAN). The meter then automatically  calculates and programs in 

the requisite  scale factor  using a true linear (y=mx + b) calculation where m = (SPAN-ZERO)/(InHi-InLo) and b = ZERO, but within 

the following parameters.
1.  Positive and negative signals may be applied, but the difference between the high and the low signal inputs must be at least 1000    

     counts or Err will be indicated.
2.  Positive and Negative values for the desired reading can be entered, but the scale factor created can not exceed the Digital  

  Display Span capability of the meter which is 12,000 counts between –1999 to 9999.
3.  The internal Signal Span is limited to 3 V DC between – 1 V DC to + 2 V DC. Any outputs from an Input Signal Conditioning  

  module that exceed these limits will cause the meter to indicate overrange regardless of the Digital Display Span scaled.
Note: Most input signal conditioners have provisions for analog calibration and scaling. If the meter’s digital scale factor is set to  

read zero with a zero input (shorted input), and to read 1000 with a 1.000 V input, any pre-calibrated signal conditioner with an  

output that does not exceed – 1 V to + 2 V, will read correctly in the meter without any further calibration.

STEP A  Enter the Calibration Mode 

1)  Press the 

P

 and   buttons at the same time.

 

  Display toggles between [CAL] and [oFF].

 

2)  Press and hold  the   or   button.

 

  Display changes from [oFF] to [on].

 

3)  Press the 

P

 button. Display toggles between [ZErO] and the  

  previous zero setting.

STEP B  Select Between Two Point Digital Calibration of Input Signal and 

Two Point Analog Output

Note: If the analog output option is not present, Step B is skipped and the 

program goes directly from Step A to Step C.

 

1)  Press the   or   button to select the display toggling from [CAL]   

  to [inPt] input calibration.

 

2)  Press the 

P

 button. Display toggles between [ZErO] and the  

  previous zero setting.

STEP C  Set the Meter’s Low Input Signal Reading on the Digital Display

 

1)  Apply a zero or low signal to the meter.  

  (Positive or negative values are allowed)

 

2)  Using the   and   buttons, adjust the meter display to the  

 

  desired reading for the applied low input signal.

 

3)  Press the 

P

 button. Display toggles between [SPAn] and the  

  previous span setting.

STEP D  Set the Meter’s High Input Signal Reading on the Digital Display

 

1)  Apply a high input signal to the meter.

 

2)  Using the   and   buttons, adjust the digital display to the  

 

  desired reading for the applied high input signal.

 

3)  Press the 

P

 button.

The Digital Calibration Procedure Mode is Now Complete

If the digital calibration was successfully completed, the menu  

 

branches to the DISPLAY FUNCTION CONFIGURATION MODE,  

 

(see page 7) and the display flashes [dP] and the previous setting.

ERROR  Indicates Unsuccessful Calibration  

 

 

If the calibration was unsuccessful, the display indicates [Err],  

 

the new calibration settings just entered will not take effect and  

 

the previously stored setting will remain. 

 

The three most likely causes of an error during calibration are:

 

1)  The full scale and zero signals were too similar. The full scale  

 

  signal must be at least 1000 counts greater than the zero or  

 

  low input signal (positive and negative values are allowed).

 

2)  The scaling requirement exceeded the capability of the meter  

 

  (–1999 to 9999).

 

3)  No input signal present, or incorrect connections.

Auto Calibration Mode

Auto Calibration Procedure

MAIN MENU

Operational Display

Sub-menu 

MODE

STEP A  

Calibration

Mode

STEP B  

Calibration

Mode

DECIMAL POINT AND 

BRIGHTNESS SELECTION

See Page 6

STEP C 

Zero

STEP D  

Span

DECIMAL POINT AND 

BRIGHTNESS SELECTION

See Page 6

Decimal Point 

(dp)

Err. Any new setting 

canceled and previous 

settings are retained

TWO POINT ANALOG 

OUTPUT CALIBRATION

SEE PAGE 6

Summary of Contents for DL-40JANUS-SG

Page 1: ...utput Range Setting and Calibration 6 General Features Specifications Software Features Index LEOPARD FAMILY DL 40JANUS SG Two scaling methods MANUAL no external sig nal source required and AUTO requi...

Page 2: ...by pressing and releasing the or buttons When there are more than two display selec tions they are shown in brackets below the first display and are also selectable by press ing and releasing the or b...

Page 3: ...L POINT AND BRIGHTNESS SELECTION See Page 6 Setpoint 3 SP3 Setpoint 4 SP4 Valley Offset Scale Factor Calibrate Analog Output Lo Calibrate Analog Output Hi X XXX XX XX XXX X XXXX XXXX 2 3 4 Decimal Poi...

Page 4: ...t STEP B Select Between Calibration of Input or Output Note If the analog output option is not present Step B is skipped and the program goes directly from Step A to Step C 1 Press the or button to se...

Page 5: ...Point Analog Output Note If the analog output option is not present Step B is skipped and the program goes directly from Step A to Step C 1 Press the or button to select the display toggling from CAL...

Page 6: ...s the P and buttons at the same time Display toggles between CAL and oFF 2 Press the P button Display shows previous dP selection STEP E Set the Decimal Point 1 Using the and adjust the display to the...

Page 7: ...unts and does not include the deci mal point If you need the decimal value with decimal point you will need to read the register convert it to decimal read 40024 see below to determine decimal locatio...

Page 8: ...ST2 Setpoint 2 Hysteresis setting 40019 0x0012 SINT16 R W 1999 9999 PRES3 Setpoint 3 setting 40020 0x0013 UINT16 R W 0 9999 HYST3 Setpoint 3 Hysteresis setting 40021 0x0014 SINT16 R W 1999 9999 PRES4...

Page 9: ...P2 Delay on Make doM Delay Time Setting STEP H Set the SP2 Delay on Break dob Delay Time Setting STEP I Set the Hysteresis Setting for Setpoint 2 1 Using the and buttons adjust the display to the desi...

Page 10: ...with external signal OFF 17 16 19 19 18 MANUAL rescaling ON AUTO scaling OFF ON SP1 NO SP1 NO SP2 NO SP2 NO Pins 14 and 15 AC DC Power Input Auto sensing AC DC power supply For voltages between 85 265...

Page 11: ...air circulation 3 Separate input output leads from power lines to protect the meter from external noise Input output leads should be routed as far away as possible from contactors control relays tran...

Page 12: ...71 1 8 DIN Cutout spacers 87 4mm 3 45 Various bezel colors are available Black is standard Prog SP3 SP4 SP2 SP1 For additional strength extra Mounting Slide Clips can be ordered and doubled up one beh...

Page 13: ...se is made from fire retardant polycar bonate A metal surround case can be ordered to enhance the meter s fire retardant capabilities and also provide shielding against electromagnetic interference EM...

Page 14: ...OA____ DL 40JANUS SG Add to the basic model number the order code suffix for each standard option required The last suffix is to indicate how many different special options and or accessories that you...

Reviews: