Hardware Manual COMBILOG 1020
128
If a bit in the module status is set it indicates that an error has oc-
curred in the datalogger. Valid is:
bit 1 = 1: EEPROM error
bit 5 = 1: (RTC-Error)
bit 2 = 1: FLASH error
bit 6 = 1: (currently not occupied)
bit 3 = 1: ADC error
…
bit 4 = 1: configuration error
bit 16 = 1:(currently not occupied)
<Kinfo> channel information
length = 29 byte
<channel type>..........................binary................................ 1 byte
<channel notation>....................ASCII .............................. 20 byte
<data format> ............................binary................................ 1 byte
<field length>.............................binary................................ 1 byte
<decimals> ................................binary................................ 1 byte
<unit> ........................................ASCII ................................ 4 byte
<host input>...............................binary................................ 1 byte
Coding <channel type>:
hex 00:
Empty Channel (EM)
hex 01:
Analog Input Channel (AI)
hex 02:
Arithmetic Channel (AR)
hex 03:
Digital Output Channel (DO)
hex 04:
Digital Input Channel (DI)
hex 05:
Setpoint Channel (SP)
hex 06:
Alarm Channel (AL)
Coding <data format>:
hex 00:
no format
hex 01:
B O O L
hex 02:
INTEGER
hex 03:
R E A L
Coding <host input>:
hex 00:
host input is not possible
hex 01:
host input is possible (tare/reset/dig.out/setpoint values)
Summary of Contents for COMBILOG 1020
Page 1: ...COMBILOG 1020 Datalogger Hardware Manual Version 3 09...
Page 2: ...Hardware Manual COMBILOG 1020 2 Issue 03 11 2005 Technical data are subject to change...
Page 4: ...Hardware Manual COMBILOG 1020 4...
Page 6: ...Hardware Manual COMBILOG 1020 6...
Page 12: ...Hardware Manual COMBILOG 1020 12...
Page 86: ...Hardware Manual COMBILOG 1020 86 Bild 8 1 Example Master Slave System...
Page 169: ...Hardware Manual COMBILOG 1020 169...
Page 170: ...Hardware Manual COMBILOG 1020 170...