read data from channel:
length = 2..5 byte
<channel number>.................... binary ................................ 1 byte
<channel value Kx> .................. binary .....................1, 2 or 4 byte
write data to a channel:
length = 2..5 byte
<channel number>.................... binary ................................ 1 byte
<channel value Px> .................. binary .....................1, 2 or 4 byte
tare/reset a channel:
length = 1 byte
<channel number>.................... binary ................................ 1 byte
read, write and tare/reset channels:
length
≥
1 byte
<tare/reset> .............................. binary ................................ 1 byte
<channel value P1> .................. binary .....................1, 2 or 4 byte
<channel value P2> .................. binary .....................1, 2 or 4 byte
:
.
.
<channel value Pn> .................. binary .....................1, 2 or 4 byte
ReqDataUnit: [<tare/reset> [ <P1> [ .... [ <Pn>] ] ] ]
ResDataUnit: <K1 > . . . . . . . . . . . . . . . . < Kn >
If a bit is set in the <tare/reset> byte, the corresponding sensor
channel is tared or reset respectively. The values following the
<tare/reset>-byte are allocated to the writeable channels of the
datalogger, according to the order of their appearance. Writeable
channels are setpoint channels and digital output channels.
Summary of Contents for COMBILOG 1020
Page 1: ...COMBILOG 1020 Datalogger Hardware Manual Version 3 09...
Page 2: ...Hardware Manual COMBILOG 1020 2 Issue 03 11 2005 Technical data are subject to change...
Page 4: ...Hardware Manual COMBILOG 1020 4...
Page 6: ...Hardware Manual COMBILOG 1020 6...
Page 12: ...Hardware Manual COMBILOG 1020 12...
Page 86: ...Hardware Manual COMBILOG 1020 86 Bild 8 1 Example Master Slave System...
Page 169: ...Hardware Manual COMBILOG 1020 169...
Page 170: ...Hardware Manual COMBILOG 1020 170...