Hardware Manual COMBILOG 1020
147
12
SPECIFICATIONS
12.1 Power Supply
Voltage range:
+10 .. 18 VDC
Power input:
depending on internal clock frequency
and scan rate
5 MHz version:
approx. 70 mW at measuring rate 60 s =
5.83 mA (at 12 V)
approx. 75 mW at measuring rate 10 s =
6.25 mA (at 12 V)
approx. 420 mW at measuring rate 1 s =
35 mA (at 12 V)
20 MHz version:
approx. 400 mW
12.2 Signal Inputs/Outputs
Analog inputs:
8, non-isolated
Digital inputs/outputs: 6 (configurable)
12.3 Signal Processing
Scan rate:
0.5, 1, 2, 3, 4, 5, 10, 20, 30 sec.
1, 2, 3, 4, 5, 10, 20, 30, 60 min.
Averaging interval:
1, 2, 3, 4, 5, 10, 15, 20, 30 sec.
1, 2, 3, 4, 5, 10, 15, 20, 30 min.
1, 2, 3, 4, 6, 8, 12h
A/D-conversion:
Delta-Sigma, precision ca.16 Bit
Accuracy of the
real time clock:
<1 s / day