Hardware Manual COMBILOG 1020
151
Remarks:
(1) Measured in 3 m distance, evaluated with 10 dB higher level limits.
(2) Influence on RS 485 communication in a range (19.66
±
3) MHz is possible.
(3) In a range 80...100 MHz with field intensities > 3 V/m and 100...115 MHz with
field intensities > 7 V/m , variation of display contrast and sporadic communica-
tion errors or interrupts, like checksum error, timeout etc., are possible.
With field intensities up to 3 V/m, the influence on the data acquisition, choosing
a sensitive range, may be as follows:
Voltage measurement
(6.25 mV range):
±
50 µV typical;
±
200 µV max.
Current measurement
(1 mA range):
±
1 µA typical,
±
5 µA max.
Resistance
measurement
(200
Ω
range):
±
0.1
Ω
typical,
±
0.2
Ω
max.
With field intensities up to 10 V/m, this influence may be as follows:
Voltage measurement
(6.25 mV range):
±
1 mV typical,
±
5 mV max.
(125
±
15) MHz
Current measurement
(1 mA range):
±
4 µA typical,
±
20 µA max.
(125
±
15) MHz
Resistance
measurement
(200
Ω
range):
±
2
Ω
typical,
±
10
Ω
max.
(125
±
15) MHz
(4) The COMBILOG datalogger is not equipped with any HF-field sensitive compo-
nents, therefore according tests are not relevant. Such tests were still accom-
plished; no influence was detected.
(5) Influence on RS 485 communication from approx. 0.4 kV: self-restoring.
(6) Influence on display from approx. 0.4 kV: self-restoring.
Refer to
: Test report COMBILOG, DOCID: BW024003.DOC dated 22.02.1996.
The requirements of EMV rules 89/336/EWG are fulfilled for domestic areas. Further
EU rules concerning CE-labelling are not applicable at present.
The instrument is equipped with CE label.
Summary of Contents for COMBILOG 1020
Page 1: ...COMBILOG 1020 Datalogger Hardware Manual Version 3 09...
Page 2: ...Hardware Manual COMBILOG 1020 2 Issue 03 11 2005 Technical data are subject to change...
Page 4: ...Hardware Manual COMBILOG 1020 4...
Page 6: ...Hardware Manual COMBILOG 1020 6...
Page 12: ...Hardware Manual COMBILOG 1020 12...
Page 86: ...Hardware Manual COMBILOG 1020 86 Bild 8 1 Example Master Slave System...
Page 169: ...Hardware Manual COMBILOG 1020 169...
Page 170: ...Hardware Manual COMBILOG 1020 170...