EN
11
Bluetooth Wireless Technology Information
Electromagnetic Compatibility (EMC)
Statement for Home Healthcare Environment
The RecoveryAir PRO system has been evaluated to international standard IEC 60601- 1-2
“General requirements for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests”.
The Medical Electrical Equipment needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided here within the
accompanying documents.
Portable and mobile RF communication can affect the Medical Electrical Equipment. See below
recommended separation distances between portable and mobile RF communication equipment
and the RecoveryAir PRO system.
Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-talkies can affect this equipment and should
be kept at least a distance d = 3.3 m away from the system.
*Note: The distance calculated from 800 MHz to 2.5 GHz
Bluetooth Compliance
Version 4.2 low energy
Operating Frequency
2.402–2.480 GHz
Output Power
0 dBm
Operating Range
3-meter radius (line of sight)
Network Topology
Star - bus
Operation
Slave
Antenna Type
Integrated chip type antenna
Modulation Type
Adaptive Frequency Hopping
Data Rate
Over the air: 1 Mbit/second
Application throughput: 0.27 Mbit/s
Data Latency
6 ms
Data Integrity
Adaptive Frequency Hopping
Robustness
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Quality of Service
This device uses Bluetooth smart technology for
wireless communication, which allows for reliable
communication in electrically noisy environments.
If connection is lost, the device will automatically
reconnect in a few seconds.
Bluetooth Profiles Supported
GAP, GATT, SM, L2CAP and Integrated Public
Profiles
Authentication and Encryption
Supported
EMC Manufacturer Declarations
Rated Maximum
Output Power of
Transmitter (W)
Separation Distance*
according to Frequency
of Transmitter (m)
0.01
0.23
0.1
0.73
1
2.3
10
7.3
100
23