Rev 1.4 09/13/16
7
Due to variations in the ambient (temp. and R.H.) zero calibration must be initially and
periodically checked to assure best performance. (See Calibration Section)
10. Locate SETUP jumpers JP-1 thru 9 on Upper Circuit Board (
Figure 2
). Select required settings
from the following menu:
JP
FUNCTION
JP-1
In house
ONLY.
JP-2
Optional AutoCalibration (See 11a)
JP-3
Set scan rate, 0-10 sec. adjustable (See 11b)
JP-4
Set gas type and channel (See 11c)
JP-5
In House Use
ONLY.
System set-up info.
JP-6
In House Use
ONLY.
Set Clock
JP-7
In House Use
ONLY
. (Contact mfg. for changes)
JP-8
Relay Latch and jumper storage. (See 11d)
JP-9
Optional Audible alarm.(See 11e)
Figure 7 - Function Jumper Settings
11. After selecting from the above, the LCD will show a new mode (SETUP, etc.) and the MAN,
ZERO and TEST push-buttons will have new functions.
Refer to Figures 2 and 5 before proceeding
a.
Auto Calibration
(Optional) select JP-2, LCD shows ACON. Haloguard
TM
compares temp.
and RH every 4 hours to compensate for changes in ambient.
b.
Scan Rate
- Rate is factory set at 4 sec. per active channel. To change, select JP-3, push
ZERO or TEST and step to desired rate. Remove jumper from JP-3.
c.
Channel and Gas Type
- See
Figure 10 - Initial Factory Setting
for channel set-up. To
change, select JP-4. Press ZERO or TEST and sequence to channel (1 thru 6) desired. Press
MAN to initiate gas type menu. Press ZERO or TEST and scroll to gas type desired (Note:
“None” selected will skip channel). Proceed to next channel and follow same process. When
finished, remove jumper from JP-4.
d.
Relay Latch Setup
- Recommended factory setting is Unlatched Mode; the alarm LEDs and
relays will return to a normal status when gas concentration goes below alarm set point. To
select Latched Mode place jumper on JP-8. Press RESET button to release alarm or check
for false alarm (Note: Alarm will be disengaged for 20 minutes).
e.
Audible Alarm
- JP-9 is factory installed to activate alarm circuit. To silence alarm remove
JP-9 jumper.
12.
TWA Integration
, (Optional). Haloguard
TM
calculates an 8 hr.
TLV-TWA,
displays this value
on LCD, activates LO alarm LED and relay if preprogrammed PPM value is exceeded.
(
Haloguard
TM
must be powered up for 8 hrs. to build database before operating).
13. Trim optional analog outputs by entering TEST mode (See
TESTING
) and turning
potentiometers on AN 1 thru AN 4 (see Figure 2 Upper Circuit Board) until output measures
5.0 or 10.0 VDC as required.
CALIBRATION
1. The Haloguard
TM
has been factory calibrated under standard conditions (70
o
F, 65% R.H.).
Due
to variations in the ambient (temp. and R.H.) zero calibration must be initially and
periodically checked to assure best performance.