3200 Series
User Manual
Part No HA028651 Issue 11.0 July-12 83
13.2.4
To Configure the Programmer
The programmer can be configured in Level 2 as explained in section 4. The Event outputs, however, can only be
configured in Level 3 or Configuration level as follows:-
Select Access Level 3 or Configuration level as described in section 6.1.3.
Operation
Action
Display View
Notes
Select the
TIMER
page
Press
as many times as
necessary to ’TIMER’
tim e r
Configure the
Timer as a
Programmer
Press
to select ‘TM.CFG’
Press
▼
or
▲
to ‘
ProG
’
prog
tm . cfg
Set the
Resolution
Press
to select ‘TM.RES’
Press
▼
or
▲
to ‘
Hour
or
‘
min
’’
Hour
Tm . res
This sets the dwell function in seconds, minutes
or hours. In this example the dwell period is set
in hours.
Set the
Threshold
Press
to select ‘THRES’
Press
▼
or
▲
to adjust
5
th res
In this example the dwell periods will not start
until the PV is within 5 units of the setpoint
Set the action
when the
programmer times
out
Press
to select ‘END.T’
Press
▼
or
▲
to ‘
Off
’ or
‘
SP2
‘ or ‘
Dwel
’
dwel
end. t
In this example the controller will continue to
control indefinitely at the last setpoint.
OFF will turn the output power off and SP2 will
control at setpoint 2
Set the
Servo
Mode
Press
(twice) to select ‘SERVO’
Press
▼
or
▲
to ‘
PV
’ or ‘
SP
‘
pV
se rvo
In this example the program will start from the
current value of the process variable
Set the first
Target
Setpoint
Press
to select ‘TSP.1’
Press
▼
or
▲
to adjust
100
Tsp. 1
In this example the setpoint will ramp from the
current value of the PV to the first target - 100
Set the first
Ramp
Rate
Press
to select ‘RMP.1’
Press
▼
or
▲
to adjust
8.0
Rm p.1
In this example the setpoint will ramp to 100 at
8.0 units per hour. The ramp units are set up in
the Setpoint List see section 10.1.
Set the first
Dwell
Press
to select ‘DWEL.1’
Press
▼
or
▲
to adjust
2:11
Dwe l.1
In this example the setpoint will dwell at 100 for 2
hours 11 minutes
Repeat the above three steps for all segments
Set the segment in
which the relay
operates
Press
to select ‘EVENT’
Press
▼
or
▲
to adjust
4
even t
Set as described in section 13.2.3.
In this example the event output will be active
during Ramp 2.
Set the number of
times the whole
program repeats
Press
to select ‘P.CYCL’
Press
▼
or
▲
to adjust
1
p. cyc l
1
= Program runs once
To
100 = Program repeats 100 times
Configure Output
4 (AA Relay) as the
Event output
Press
to select ‘AA’ List
Press
to select ‘4.SRC.A’
Press
▼
or
▲
to select ‘
PrG.E
’
Pr.ge
4. src. a
This can only be done in Configuration level.
You can also select 4.SRC.B, 4.SRC.C, or 4.SRC.D
or assign these to other functions, for example
‘
t.run
’ or ‘
t.End
’ so that the relay also operates
when the timer is running or when it ends.
Note, however, if the same output is used to
operate when the program is running it cannot
also operate during a segment. In this case use
an alternative output is described in section
13.2.2.
Summary of Contents for XST-2-0-12-1V1-E28
Page 16: ...12 FURNACE TROUBLE SHOOTING ...
Page 27: ...ɸ ...
Page 28: ...ɸ ...
Page 29: ......
Page 30: ......
Page 31: ...3216 3208 3204 User Manual 3200 Range Process Controller HA028651 11 Aug 2012 ...
Page 170: ...User Manual 3200 Series 138 Part No HA028651 Issue 11 0 July 12 21 Certificate of Conformity ...
Page 171: ......
Page 173: ...Model 2416 Process Controller User Manual Part No HA025041_12 Date August 2014 ...
Page 174: ...This page is intentionally blank ...
Page 259: ...This page is intentionally blank ...