SPECIFICATIONS
High Voltage Power Supplies
(refer to page 2 for models)
Type Output:
Constant Voltage, Constant Watts or
Constant Milliamps with automatic crossover
Maximum Voltage:
2000 – 6000 Volts, (depending on model)
Maximum Current:
200 – 600 Milliamps (depending on model)
Maximum Power:
200 – 300 Watts (depending on model)
Regulation:
≤
1%
Accuracy:
± 1.5% full scale for each display
Number of
Output Terminals:
Four recessed sets of 4 mm sockets
Safety Interlock:
Load sensing shut-down-on-disconnect. D.C.Output On
key actuation necessary to begin voltage generation.
In the event of shutdown due to power interruption,
automatic restart is provided.
Timer:
00 to 99 hrs. 59 min.
Ground Leakage:
Leakage of 400 microamps or more will interrupt the
generation of high voltage.
Input Power:
115 VAC/60 Hz./400 W
230 VAC/50 Hz./400 W
Ambient Operating
Temperature Range:
0 ° – 30 °C (non-condensing atmosphere)
Dimensions:
11.25
" (D) x 12.6" (W) x 5" (H)
28.6 cm x 32 cm
x 12.7 cm
Weight: net:
15.4 lbs. 7 kgs.
shipping:
17 lbs.
7.7 kgs.
14
BASIC OPERATING
INSTRUCTIONS
Operation in
Constant Voltage:
1.
The electrophoresis chamber
should first be set up, then filled
with buffer and sample, before
plugging the chamber’s leads
into the appropriate positive
and negative 4 mm sockets
(14) on the power supply.
2.
Switch on the power supply
using the Power On/Off
switch (22). The main power
switch is marked with a 1 to
indicate on and a 0 to indicate
off. Once on, the Green D.C.
Output Off light (21) and
Yellow Actual Display light (17)
will be illuminated. The Volts
display (3), Milliamps display
(5) and Watts display (7) will
all indicate zero. The Time dis-
play (10) will indicate zero as
well.
3.
Gently press the Set Display
key (19). The actual switch itself
is located in the grey area
located immediately below the
word “Set” in the Set Display
key. When pressed, there will
be a tactile sensation, the
power supply will beep and
the Yellow Set Display (18) light
will be illuminated. Each key on
Set
Actual
On
Off
14
22, 21,
17, 3,
5, 7, 10
19, 18
Set
Actual
Locator
Number
Light Off
Light On
DC Output
Display
Display
NOTE:
An alterntive procedure favored by some users is as follows:
Sample is withheld from the chamber until the power supply has been properly
adjusted and successfully started. Once the user has confirmed that the chamber and
power supply are both functional, the power supply is turned off and sample is
loaded. The power supply will remember its settings, so it can be turned on and start-
ed without further adjustment.
(See fold out locator on last page)
BASIC OPERATING INSTRUCTIONS
3