background image

C. Ground Leakage Detected: 

A continuous tone accompanied by the

illumination of the ground leakage light. The voltage and wattage displays
will hold their last value. The milliamps display will go blank. (See below
“Ground Leakage”)

D. Power On Restart:

Nine short bursts in groups of three each with

each output display showing decimal points. This alarm activates to alert
the operator to the fact that auto-restart will occur at the end of a mains
power failure. The alarm will clear itself.

E. Overheating Detected: 

A continuous tone accompanied by flashing

displays. This alarm indicates the heatsink, located at the rear of the unit,
has exceeded safe operating temperatures. Momentarily turning off the
power supply will clear the alarm. (See Service)

F. Mains Overload:

A continuous tone accompanied by flashing displays.

This alarm is indistinguishable from the “Overheating detected” alarm. It
indicates an internal fault or a temporary surge in the mains supply.
Momentarily turning off the power supply will clear the alarm. (See Service)

Unit Will Not Start: 

This power supply is equipped with a load sensing

interlock which is designed to inhibit the generation of high voltage whenever
an open connection is detected at the output.

Open Connection: 

An open connection is defined as a break in the path

in which electricity normally flows. This could be caused by any of the follow-
ing conditions:

1.

A broken electrode within the electrophoretic chamber.

2.

A broken wire within the connecting cord (either positive or negative).

3.

Insufficient buffer levels within the electrophoretic chamber.

4.

Loose connectors in any electrical connection going to or coming from the
power supply or chamber.

5.

Electrochemical energy stored within the chamber. (See below, “Unit Will
Not Restart ”)

Unit Will Not Restart:

Certain types of electrophoretic techniques will,

over a period of time, store energy within the electrophoretic chamber. This
charge may inhibit the normal function of the load sensing interlock when
attempting to restart an experiment which has been temporarily interrupted.
To overcome this effect depress and hold the D.C. Output On key until the
power supply reaches its proper operating voltage. Upon releasing the D.C.
On key the power supply should remain engaged. If it will not remain
engaged refer to the section entitled “Unit Will Not Start”.

No Displays Light:

This power supply is equipped with 2 circuit breakers

located at the rear of the instrument. If either circuit breaker is activated the
power supply will not turn on. When activated, the circuit breakers expose a
white stem.

12

6.

Gently press the D.C. Output
On key (24). The Red D.C.
Output On light (23) should illu-
minate and the D.C. output as
indicated by the three displays
(3, 5, 7) should rapidly
increase until the operating volt-
age limit is reached. At this
point the Yellow At Limit light
(volts) (2) should be the only “at
limit” light illuminated.

On

Off

DC Output

LIMITS

Electrophoretic chambers are generally designed for a relatively specific purpose.
For example, horizontal chambers use agarose gels to separate DNA or RNA frag-
ments while a DNA Sequencing chamber is almost always used to separate DNA
in a denaturing polyacrylamide gel. In each case, the voltage, milliamp and
wattage requirements are well– defined within a reasonable range of values. In
cases such as these, the user can safely assume that the manufacturer has designed
the chamber to withstand the voltage and heat energy necessary to perform the
electrophoretic separation when standard protocols are followed.
Some types of electrophoretic chambers are specifically designed to be multipur-
pose devices. For example, a vertical slab gel chamber could be used for anything
from DNA Sequencing to Isoelectric Focusing depending on the gel type and buffer
system used. Choosing safe operating limits for a chamber of this type requires a

24, 23,
3, 5, 
7, 2

5

NOTE:  

The power supply output may be adjusted while operating. If you turn any

of the three adjustment knobs while the D.C. Output On light is illuminated, the
power supply will automatically switch to the Set Display Mode. Each display will
show the set operating limit while in this mode and new limits may be set. One sec-
ond after releasing the adjustment knob the power supply will revert to the Display
Actual Mode.

Operation in Constant Current: 

The method used to set the power supply

in constant current is the same as the method used for constant voltage. The only
difference involves the selection of an operating limit for current which, as the out-
put increases, is attained before the operating limits for voltage or watts.

Operation in Constant Watts: 

The method used to set the power supply in

constant watts is the same as the method used for constant voltage. The only dif-
ference involves the selection of an operating limit for watts which, as the output
increases, is attained before the operating limits for voltage or current.

THERMO EC HIGH VOLTAGE POWER SUPPLIES

THERMO EC HIGH VOLTAGE POWER SUPPLIES

Summary of Contents for EC2060

Page 1: ...nstructions carefully before using this power supply A VERTISSEMENT Veuillez avant tout emploi du générateur lire attentivement ce manuel d utilisation 159 3037 00 Rev A 02 01 Printed in USA 100 Colin Drive Holbrook NY 11741 4306 Tel 1 631 244 2929 Toll Free 1 800 EC RANGE FAX 1 631 244 0606 For Technical Service 1 800 327 2643 Email savantec savec com www thermoec com A Thermo Electron business ...

Page 2: ...ENT Ce générateur a été conçu pour être utilisé comme source de courant DC pour l électrophorèse et il est capable de générer un courant mortel Prenez les mêmes précautions que pour tout autre appareil électrique N utilisez pas l appareil sans que le couvercle de la chambre soit placé Ne raccordez pas les sorties à la terre N utilisez pas l appareil dans des environnements humides où la condensati...

Page 3: ...hip Between Volts Milliamps Watts and Chamber Resistance 15 Appendix B Running Multiple Chambers 15 Guarantee 16 Locator Guide fold out back page 1 G U A R A N T E E This laboratory equipment was produced by Thermo EC with the highest practical standards of materials workmanship and design The design and manufacture of parts have been conceived with one purpose to produce a unit which will give sa...

Page 4: ... change For added convenience this power supply will restart itself in the event of a power failure This will not reset the operation of the count down timer 2 APPENDIX A Relationships Between Volts Milliamps Watts and Chamber Resistance There are three fundamental concepts which form the basis for understanding the relationship between volts milliamps and chamber resistance When combined with the...

Page 5: ...tive and negative 4 mm sockets 14 on the power supply 2 Switch on the power supply using the Power On Off switch 22 The main power switch is marked with a 1 to indicate on and a 0 to indicate off Once on the Green D C Output Off light 21 and Yellow Actual Display light 17 will be illuminated The Volts display 3 Milliamps display 5 and Watts display 7 will all indicate zero The Time dis play 10 wil...

Page 6: ...g limits has been included to guide the user in selecting proper operating limits This section details both the method and rationale used in selecting these operating limits Volts 1 3 NOTE Each adjustment knob has a RAPID ADJUSTMENT MODE which is activated whenever the knob is rotated rapidly 4 WARNING This power supply is not equipped with any user serviceable parts Erratic Operation Unstable err...

Page 7: ...cuit breakers located at the rear of the instrument If either circuit breaker is activated the power supply will not turn on When activated the circuit breakers expose a white stem 12 6 Gently press the D C Output On key 24 The Red D C Output On light 23 should illu minate and the D C output as indicated by the three displays 3 5 7 should rapidly increase until the operating volt age limit is reac...

Page 8: ...erat ing temperature the chambers will withstand is a critical aspect of safe operation Most electrophoretic chambers with a few notable exceptions are made of acrylic plastic and must operate well below a temperature of 55 C If there is a lack of information about the capabilities of the chamber being used regular monitoring of the operating temperature is recommended This should be done without ...

Page 9: ...ht 11 will be illuminated and the timer will begin counting down Resetting 1 Switch on the power supply using the Power On Off switch 22 Once on the Green D C Output Off light 21 and the Yellow Actual Display light 17 will be illuminated The Volts display 3 Milliamps display 5 and Watts display 7 will all indicate zero 2 Gently depress the center of the Arrow Down key 12 When pressed there will be...

Page 10: ...ge As the voltage increases during the experment it will eventually reach the operating limit for voltage and automatically crossover to constant voltage operation At this point the current will start to drop as the buffer breaks down but the experment and chamber will be unharmed 9 6 Gently press the D C Output On key 24 The Red D C Output On light 23 should illuminate and the D C output as indic...

Reviews: