Chapter 7 Servicing
7-14
Fullscale Analog Outputs Screen
6. Repeat the above steps to ensure the adjustments are accurate.
7. Re-install the instrument cover.
PRESSURE TRANSDUCER ADJUSTMENT
This procedure should only be performed by an instrument service technician.
Equipment Required:
Vacuum
Pump
Screwdriver
CAUTION:
Some internal components can be damaged by small amounts of static
electricity. A properly grounded antistatic wrist strap must be worn while handling any
internal component.
1. Wear an antistatic wrist strap that is properly connected to earth ground, see “Safety
Precautions,” earlier in this chapter for more information.
2. Remove the instrument cover.
3. Disconnect the tubing from the pressure transducer and connect a vacuum pump
known to produce a vacuum less than 1 mm Hg.
4. From the Run screen, press
MENU
to display the Main Menu. Use the
↓
pushbutton
to move the cursor to Service menu. Press
ENTER
to display the Service menu. Use
the
↓
pushbutton to move the cursor to Pressure. Press
ENTER
to display the
pressure reading.
5. Adjust the zero potentiometer on the pressure transducer for a reading of zero mm
Hg.
6. Disconnect the vacuum pump. The display should read the current local barometric
pressure. If this value does not agree with a known accurate barometer, adjust the
span potentiometer.
7. Reconnect the tubing to the pressure transducer.
8. Re-install the instrument cover.
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......