Chapter 2 Installation
Analog Output Cover Installation
This analog output cover must be mounted over the analog outputs to comply with
89/336/EEC Directive. This procedure describes how to install the user-supplied analog
output cable in the instrument’s analog output cover. The following shielded cables or
their equivalent are recommended:
Cable
Gauge
No. of Conductors
Alpha
#1741C
20
2
Alpha
#1746C
18
2
Alpha
#5320/2C
*
20
2
Alpha #51 52C
*
20
2
Alpha
#5162C
*
18
2
Alpha
#1743C
20
4
Alpha
1747/4C
18
4
Alpha
#5320/4C
20
4
Alpha
#5154C
20
4
Alpha
#5164C
18
4
Belden
#8208
18
2
* Maximum shielding. Under harsh environments, maximum shielding may be required.
The following tools are required:
Small
screwdriver
Wire stripper
Electrical tape or heat shrink tubing
The user-supplied shielded analog output cable must be properly grounded
by coming
into full contact with the cable clamp (mounted to the analog output cover). To ensure
full contact, the shielding must be exposed and folded back over the cable as shown in
Figure 2-3.
Figure 2-3.
Shielded Cable with Shielding Pulled Back
2-4
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......