Chapter 3 Operation
3-51
Min and Max Chamber Temperature Limits
. The Min Chamber
Temperature limit screen, shown below, is accessible only when the instrument is in the
service mode. It is used to change the min chamber temperature alarm limit. The min
and max Chamber Temperature screens function the same way.
To display the Min or Max Chamber Temperature limit screen (service mode on):
From the Main Menu choose Alarm
From the Alarm menu choose Chamber Temperature
From the Chamber Temperature menu choose Min or Max
To use the Min or Max Chamber Temperature limit screen (service mode on):
Use the
↑
and
↓
pushbuttons to increment/decrement the value
Press
ENTER
to accept the change
Press
MENU
to return to the Chamber Temperature menu
Press
RUN
to return to the Run screen
Set Min Chamber Temperature Screen
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......