Initial S/R Ratio
The Initial S/R Ratio screen, shown below, displays both the initial S/R ratio and the
current S/R ratio. The initial S/R ratio is determined at the factory, and is used to correct
for the slight variations found from one correlation wheel to another. The only time the
initial S/R ratio should be changed is when the correlation wheel is replaced or after
factory consultation. Normally, when zero air is being sampled, the sample/reference
ratio is between 1.14 and 1.18.
To display the Initial S/R Ratio:
From the Main Menu choose Service Mode
From the Service Mode menu choose Initial S/R Ratio
To use the Initial S/R Ratio screen:
Use the
↑
and
↓
pushbuttons to increment/decrement the value
Use the
←
and
→
pushbuttons to move the cursor left and right
Press
ENTER
to accept the change
Press
MENU
to return to the Service Mode menu
Press
RUN
to return to the Run screen
Initial S/R Ratio
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......