Chapter 3 Operation
3-71
Hi or Lo Multi-Point Calibration
Up to three gas concentrations (cal-points) for each range may be calibrated to using the
following steps. Three cal-points will give the most accurate readings over the entire
range. The Calibration process is sequential and will only work properly if all steps are
followed in order. The examples below use the Lo Multi-Point Cal screens, but the
procedure is the same for the Hi Multi-Point Cal screens.
To view the Hi or Lo Multi-Point Cal Screen:
From the Main Menu choose Service Mode
From Service Mode choose Hi or Lo Multi-Point Cal
COEFFICIENTS
DEFAULT COEF
Lo Multi-Point Cal Screen
LO MULTI-POINT CAL:
>CAL POINT 1
CAL POINT 2
CAL POINT 3
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......