Section 2 - Calibrate
2.1 Calibrate Mode
Once the freezer has stabilized, the control probe may need
to be calibrated. Calibration frequency is dependent on use,
ambient conditions and accuracy required. A good laboratory
practice would require at least an annual calibration check. On
new installations, all parameters should be checked after the
stabilization period.
Before making any calibration or adjustments to
the unit, it is imperative that all reference instru-
ments be properly calibrated.
a. Calibrating the Control Probe
Plug a type T thermocouple reader into the receptacle
located inside the lower door (see Figure 1-5). Compare the
control temperature set point to the temperature of the measur-
ing device.
1. Press the Mode key until the Calibrate indicator lights.
2. Press up/down arrow to match the display to calibrated
instrument.
3. Press Enter to store calibration.
4. Press the Mode key to return to Run mode.
Temperature Stabilization Periods
Startup
- Allow 12 hours for the temperature in the cabinet to
stabilize before proceeding.
Already Operating
- Allow at least 2 hours after the display
reaches set point for temperature to stabilize before proceeding.
During calibration, the temperature display will not be avail-
able.
If no keys are pressed for approximately five minutes while
in calibration mode, the system will reset to Run mode.
Model 900 Series _________________________________________________________________________Calibration
2 - 1