In OMNIC Help Topics find “spectrum” in the Index and go to
“Collecting a sample spectrum.”
How do I know my spectra are meeting quality
standards?
OMNIC continuously monitors the quality of the data you collect,
based on the parameters you set. The Quality tab in the Experiment
Setup dialog box contains parameters for specifying the spectral
quality characteristics that you want checked when you collect
spectra.
OMNIC offers four categories of spectral quality checks:
•
Spectrum
checks
•
Parameter
checks
•
Background
checks
•
Interferogram
checks
When OMNIC performs a check and detects a problem, the Collect
Status indicator is displayed as a yellow circle or a red X. Click the
indicator (or click the View Collect Status button at the end of data
collection) to see a summary of data collection problems
encountered during the collection and other information about the
collection.
When a collected spectrum passes all of the selected quality checks,
the Collect Status indicator appears as a green check mark.
?
In OMNIC Help Topics find “quality checks” in the Index and go to
“Quality checks.”