4. To run the diagnostic tests for a component, click that
component’s indicator icon.
A dialog box appears showing the status of the component. If the
operating values for the component are within proper limits, a
check mark appears in the Status column. If the value is outside
the acceptable range, an X appears in the Status column.
5. When you are finished viewing the information, choose OK.
Maintaining
your spectrometer
This section describes how to use the OMNIC software to align the
spectrometer and accessories and run performance tests.
View “Maintaining your spectrometer” in Spectrometer Help Topics
for instructions for other common maintenance procedures, such as
changing desiccant, cleaning the purge filter and cleaning the outside
of the spectrometer.
Aligning the spectrometer
To ensure optimum system performance (high signal intensity, low
noise and good sensitivity), you should align the spectrometer at
least once a week using the OMNIC software. See the answer to the
question “When should I align my spectrometer” in the preceding
chapter for more information.
Note
If you have just turned on or moved the spectrometer, wait
15 minutes (one hour for best results) before you align it so that it
will have time to stabilize.
S