If your sample holder is not installed on a baseplate, use a flat-blade
screwdriver and the slotted screws to attach the holder to a baseplate
as shown below. Then install the baseplate in the sample
compartment.
If there is a sample in the
holder, remove the sample.
To install a sample or filter, slide the sample or filter into one pair of
slots. If the accessory support is currently installed in the slot you
want to use, remove the support before installing the sample or filter
(see the first illustration in this section). Replace the support when
you are finished using the slot.
Do not discard the support; it is
needed for installing slide-mounted accessories.
Nicolet FT-IR User’s Guide 29