6. Look through the sliding door to make sure there is no
sample in the sample holder and then press the OK/Enter
button to start data collection.
If there is a sample in the sample holder, remove it before
pressing the button:
If your spectrometer is purged but does not have the Smart
Purge option...
Open the sliding door, remove the sample, close the sliding
door and then wait 1 minute before choosing OK.
If your spectrometer has the Smart Purge option...
Open the sliding door, remove the sample and then slide the
door until it is open about 1 cm. When you hear the purge
flow rate return to normal, close the door all the way and
choose OK.
If your spectrometer is sealed and desiccated...
Open the sliding door, remove the sample, close the sliding
door and then choose OK.
A background spectrum appears in the Collect Sample
window. The spectrum is updated as more data are collected.
When all the data have been collected, a message appears
asking you to prepare to collect the sample spectrum:
Nicolet FT-IR User’s Guide 33