In OMNIC Help Topics find “filter wheel” in the Index and go to
“Controlling the filter wheel” for more detailed instructions.
Using the
energy screen wheel
The optional energy screen wheel lets you automatically insert an
energy screen into the beam path to block out a portion of the energy
at all frequencies of the infrared beam. This is useful for preventing
detector saturation and signal distortion. For more information about
using energy screens, see “When to use energy screens” in the
“Spectrometer Basics” chapter.
The wheel can be moved to any of four positions: an open position
(for collecting spectra without a screen in the beam path) and 3
screen positions. As shipped from the factory, the screen positions
are populated with 30%, 10%, and 3% screens. The Screen Wheel
parameters in Experiment Setup to describe the percentage of light
transmitted by each screen installed in the wheel. Simply select one
to move the screen into the beam path.
Nicolet FT-IR User’s Guide 75