Troubleshooting Software Problems
The OMNIC software automatically and continuously checks the
status of your system. If a problem is found, a message automatically
appears giving you access to more information. This chapter lists
some simple troubleshooting measures you can take to solve
software problems with OMNIC. If the action does not solve the
problem, call your Thermo Electron service office.
?
In OMNIC Help Topics find “troubleshooting” in the Index and go
to “Troubleshooting” for more information.
Problem Possible
Cause Solution
You cannot add a
spectrum to a user
library.
The resolution of the spectrum
is lower (higher numerical
value) than that of the library.
The spectrum is from a
commercial library.
Collect the spectrum at the same resolution
as the library (first set Resolution on the
Bench tab of the Experiment Setup dialog
box) and then add it to the library.
Only spectra you collect can be added to a
user library. Commercial libraries cannot be
altered and the spectra contained in them
cannot be copied to other libraries.
The Quantify
command is dimmed.
No quantitative analysis
method is selected.
No spectrum is selected.
The quantitative analysis
method is not appropriate for
the selected spectrum.
Select a quantitative analysis method using
Quant Setup in the Analyze menu.
Select a single spectrum.
Select an appropriate method for the
selected spectrum.
Nicolet FT-IR User’s Guide 135