15
ES Series Lab Freezer
High Alarm Section
Low Alarm Section
Function
Press
Hold button (s) down for
Example screen
To change the
High Alarm set point
5 seconds
Followed by
6 times
Followed by
1 second
Followed by
(Increase/Decrease)
Until required high alarm
set point
Followed by
1 second
Followed by
5 seconds to go back to
standard display
Table 8. High Alarm Section
Function
Press
Hold button(s) down for
Example screen
To change the
Low Alarm set point
5 seconds
Followed by
7 times
Followed by
1 second
Followed by
(Increase/Decrease)
Until required low alarm
set point
Followed by
1 second
Followed by
5 seconds to go back to
standard display
Table 9. Low Alarm Section