| Main Controller: 1x8 & 5x16 Segment Programmable
6-44
|
330985H01
1100°C Box Furnace
Chapter 6
End Type parameter
The action which occurs at the end of program or in reset depends on the configuration of
the
‘END.T’
parameter. The
‘END.T’
can be:
OFF : The heating is turned OFF,
dwEll : Controls at last program setpoint,
SP2 : Controls at setpoint 2 (When the timer completes the target setpoint will switch to
setpoint 2. The setpoint 2 may be a lower or a higher temperature.),
rES : Reset on completion and reverts to SP1 or SP2, based on the Setpoint selection.
SCROLL
through parameters in level 2 and set the required
‘END.T’
by pressing
or
.
OR
OR
OR
To Configure the Programmer
1. Enter level 2 to configure the timer as a programmer. Press SCROLL
button to
scroll through the list of parameters until
‘TM.CFG’
is displayed. Now, press
DOWN
or UP
button to select
‘PROG’
.
2. To set the resolution, press SCROLL
button to select
‘TM.RES’
. Press DOWN
or UP
button to set ‘Hour or ‘min’ (In this example, the ramp rate and dwell
period are set in hours).