Normal Controller Operation
Thermo Scientific
6-3
Table 3. Push button Keypad
In
14
Input type (J thermocouple in °C;°F=44
SPH
1200
High setpoint limit °C; °F =2192
SPL
0.0
Low setpoint limit °C; °F =32
Upr
oFF
Up ramp rate (degrees C or F / minute)
Dnr
oFF
Down ramp rate (degrees C or F / minute)
tmU
1
Ramp Times Units,°Per minute ( 0=°Per Hour)
AL1
9
Alarm 1 type
AL2
oFF
Alarm 2 type
HY1
1
Alarm 1 hysteresis (°F=1)
HY2
0
Alarm 2 hysteresis
SC
on
Super Control
dr
0
Direct/Reverse action
Button
Function
Pressing and holding the SET/ENT for three second advances the dis-
play to the Operation Parameters Menu.
While in the Operation Parameters Menu, use SET/ENT to move from
one parameter to next, and to register changes you have made in set-
point and parameter value.
Holding SET/ENT for three seconds exits either the Operation or Setup
Parameters menu.
Use the Up Arrow button to increase the temperature setpoint display
and change parameter values in the Operation and Setup Parameter
menus. Whenever you change the value of setpoint or parameter, the
decimal point flashes to remind you to register the changed value with
SET/ENT.
Use the Down Arrow button to decrease the temperature setpoint dis-
play and to change parameter values in the Operation and Setup
Parameter menus. Whenever you change the value of a setpoint or
parameter, the decimal point flashes to remind you to register the
changed value with SET/ENT.
SET/ENT
Summary of Contents for BF51841BC
Page 1: ......
Page 6: ...List of Figures ii Thermo Scientific ...
Page 18: ...Safety Notes Standards and Directives 1 12 Thermo Scientific ...
Page 22: ...Unpacking Unpacking 3 2 Thermo Scientific ...
Page 28: ...Installation Shelf Installation in Box Furnaces 4 6 Thermo Scientific ...
Page 38: ...Correction of Temperature Overshoot 6 8 Thermo Scientific ...
Page 48: ...Communication Option Addresses for Multiple Controllers 8 6 Thermo Scientific ...
Page 57: ...Maintenance Right Hand Door Conversion Thermo Scientific 9 9 Figure 12 Door Hinge Sheet 2 2 ...
Page 63: ...Replacement Parts Thermo Scientific 11 3 Figure 13 Wiring Diagram ...
Page 64: ...Replacement Parts 11 4 Thermo Scientific ...
Page 65: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Page 66: ...Error Codes 12 2 Thermo Scientific ...
Page 68: ...WEEE Compliance 13 2 Thermo Scientific ...
Page 69: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Page 70: ...Spare Parts and Accessories 14 2 Thermo Scientific ...