9
8 Segment & 4x16 Segment Programmable Models w/OTP
Alarms
Thermo Scientific
Large Tabletop Muffle
35
tYPE
: Press the UP or DOWN ARROW to select the segment type [End (end of program), rmP.r =
ramp rate (ramp to a specified setpoint at a set rate), rmp.t = ramp time (ramp to a specified
temperature in a set time), dwEll (to maintain a constant temperature for a set time), StEP (climb
instantaneously from current to specified temperature), cALL (to call a program as a subroutine,
available only on 4x16 programmable models)]. The controller will default to “End.” Other parameters
used with tYPE include; tGt target setpoint), Rate (rate of temperature increase) and dur (time to
target setpoint or time to dwell).
End.t
: End segment type: dwELL (dwell continuous), rSEt (reset) and S OP (End Segment Output
power level.
AL LiSt (Alarm List)
IdHi
: Deviation High Alarm.
Atun LiSt
: (Autotune List)
tunE
: One-shot autotune enable.
drA
: Adaptive tune enable.
drA.t
: Adaptive tune trigger level in display units. Range = 1 9999.
Pid LiSt
G.SP (Gain Setpoint)
: Is the temperature at which the controller switches from the (SEt1) PID values
to the (SEt 2) PID values.
Pb
: Proportional band in display units (SEt 1).
ti
: Integral time in seconds (SEt 1).
td
: Derivative time in seconds (SEt 1).
Pb2
: Proportional band (SEt 2).
ti2
: Integral time in seconds (SEt 2).
td2
: Derivative time in seconds (SEt 2).
ACCS LiSt (Access List)
Access Code (Code needed to enter or change the other configuration parameters which are not
normally accessible.) Not accessible.
Alarms
The controller will flash an alarm message in the home display if an alarm condition is detected.
IdHi
: PV deviation high alarm.
2FSH
: PV full scale high alarm.
LCr
: load current low alarm.
HCr
: load current high alarm.
Summary of Contents for F30400
Page 6: ...Contents 4 Large Tabletop Muffle Thermo Scientific...
Page 10: ...1 Safety Information Warning 8 Large Tabletop Muffle Thermo Scientific...
Page 14: ...2 Introduction Principles of Operation 12 Large Tabletop Muffle Thermo Scientific...
Page 18: ...4 Unpacking 16 Large Tabletop Muffle Thermo Scientific...
Page 28: ...7 Single Setpoint Models w OTP Tuning 26 Large Tabletop Muffle Thermo Scientific...
Page 50: ...11 Furnace Loading 48 Large Tabletop Muffle Thermo Scientific...
Page 62: ...15 Replacement Parts List 60 Large Tabletop Muffle Thermo Scientific...
Page 64: ...16 Ordering Procedures 62 Large Tabletop Muffle Thermo Scientific...