47
11
Furnace Loading
Do not overload your furnace chamber. If the load is to be heated uniformly, it
should not occupy more than two-thirds of the furnace chamber. Failure to observe
this caution could result in damage to furnace components.
For best results, use only the center two-thirds of the furnace chamber.
Failure to connect the exhaust port to an appropriate exhaust system will result in
smoke and gases filling the work area.Without the connection, gases and smoke will
escape around the door seal and at the rear of the furnace.
• Use Hearth plate to elevate load when placing on bottom heating element. This prevents bottom
heating element from overheating and burning out. (Part Numbers PHX1 & PHX2)
• If you are heating a number of small parts, spread them throughout the center of the furnace
chamber.
• Keep objects away from thermocouple.
• Use insulated tongs and mittens when loading and unloading furnace.
• Always wear safety glasses.