4-2
8600 Series
Thermo Scientific
Section 4
Alarms
This condition is created when the high stage compressor and fans run for
30 minutes and are not capable of cooling the interstage heat exchanger to
the proper temperature. Under this condition, the high stage compressor
and fans will turn off after 30 minutes and an audible and visual alarm will
occur along with the "HS SYST FAIL" message in the LED message
center.
When multiple alarm conditions occur, active messages are displayed in the
message center one at a time, updating at 5 second intervals. Pressing
Silence during multiple alarms causes all active alarms to be silenced and to
ring back in 15 minutes.
An internal communication failure has occurred with the micro board.
During this alarm, the compressor(s) attempt to run continuously.
However, with this type of failure, freezer operation becomes
undependable.
Communication between the micro board and the display board has been
lost. Under this condition, the visual alarm flashes along with dashes in the
temperature display (----). Contact Technical Services.
High Stage System Failure
Alarm (-86C units only)
Multiple Alarms
Micro Board Failure
Alarm
Lost Communication
Summary of Contents for Forma 8600 Series
Page 55: ...8600 Series 8 1 Thermo Scientific Section 8 Spare Parts...
Page 75: ...8600 Series 9 1 Thermo Scientific Section 9 Refrigeration Schematics...
Page 76: ...9 2 8600 Series Thermo Scientific Section 9 Refrigeration Schematics...
Page 77: ...8600 Series 9 3 Thermo Scientific Section 9 Refrigeration Schematics...
Page 78: ...9 4 8600 Series Thermo Scientific Section 9 Refrigeration Schematics...
Page 79: ...8600 Series 9 5 Thermo Scientific Section 9 Refrigeration Schematics...
Page 80: ...9 6 8600 Series Thermo Scientific Section 9 Refrigeration Schematics...
Page 81: ...8600 Series 10 1 Thermo Scientific Section 10 Electrical Schematics...
Page 82: ...10 2 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 83: ...8600 Series 10 3 Thermo Scientific Section 10 Electrical Schematics...
Page 84: ...10 4 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 85: ...8600 Series 10 5 Thermo Scientific Section 10 Electrical Schematics...
Page 86: ...10 6 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 87: ...8600 Series 10 7 Thermo Scientific Section 10 Electrical Schematics...
Page 88: ...10 8 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 89: ...8600 Series 10 9 Thermo Scientific Section 10 Electrical Schematics...
Page 90: ...10 10 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 91: ...8600 Series 10 11 Thermo Scientific Section 10 Electrical Schematics...
Page 92: ...10 12 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 93: ...8600 Series 10 13 Thermo Scientific Section 10 Electrical Schematics...
Page 94: ...10 14 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 95: ...8600 Series 10 15 Thermo Scientific Section 10 Electrical Schematics...
Page 96: ...10 16 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 97: ...8600 Series 10 17 Thermo Scientific Section 10 Electrical Schematics...
Page 98: ...10 18 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 99: ...8600 Series 10 19 Thermo Scientific Section 10 Electrical Schematics...
Page 100: ...10 20 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 101: ...8600 Series 10 21 Thermo Scientific Section 10 Electrical Schematics...
Page 102: ...10 22 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 103: ...8600 Series 10 23 Thermo Scientific Section 10 Electrical Schematics...
Page 104: ...10 24 8600 Series Thermo Scientific Section 10 Electrical Schematics...