Section 7
Specifications
Certifications
Declaration of Conformity is available upon request.
Safety Specifications
Indoor Use Only
Altitude - up to 2,000 meters
Temperature - 5°C to 43°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
Installation Category II
1
Pollution Degree 2
2
Class of Equipment I
Intended Use
This product is intended for use as a General Purpose Laboratory Freezer for storing samples or inventory between -50 and -86°C for units
rated at -86°C and between -10 and -40°C for units rated at -40°C.
This unit is not intended for use in an explosive environment, nor to be used for the storage of flammable inventory. This unit is not
intended for use in a Class II medical application as defined by Title 21 of the Federal Code of Regulations.
Fluorinated Greenhouse Gases
Compliant with REGULATION (EU) No 517/2014 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL on fluorinated greenhouse gases.
This product contains fluorinated greenhouse gases in a hermetically sealed system. This product contains foam blown with fluorinated
greenhouse gas, R-245fa. If a leak in the sealed system is detected, the operator shall repair without undue delay
.
For ULT models rated 230V/50Hz, intended for import into the EU, the following sizes are designed for use with the following amounts of
fluorinated greenhouse gases:
______________________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand safely. It
depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the category used for
instruments in installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial
laboratories, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2 Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2 assumes that
normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity caused by condensation.
8600 Series
7-5
Thermo Scientific
1st Stage
2nd Stage
Size (ft³)
Refrigerant
Amount (kg)
GWP
Refrigerant
Amount (kg)
GWP
3
R-404a
0.227
3 922
R-508b
0.173
13 396
12.7
R-404a
0.680
3 922
R-508b
0.425
13 396
17
R-404a
0.680
3 922
R-508b
0.439
13 396
20
R-404a
0.680
3 922
R-508b
0.454
13 396
For -86°C rated units
1st Stage
Size (ft³)
Refrigerant
Amount (kg)
GWP
3
R-404a
0.425
3 922
12.7
R-404a
0.539
3 922
17
R-404a
0.595
3 922
20
R-404a
0.652
3 922
For -40°C rated units
Summary of Contents for Forma 8600 Series
Page 55: ...8600 Series 8 1 Thermo Scientific Section 8 Spare Parts...
Page 75: ...8600 Series 9 1 Thermo Scientific Section 9 Refrigeration Schematics...
Page 76: ...9 2 8600 Series Thermo Scientific Section 9 Refrigeration Schematics...
Page 77: ...8600 Series 9 3 Thermo Scientific Section 9 Refrigeration Schematics...
Page 78: ...9 4 8600 Series Thermo Scientific Section 9 Refrigeration Schematics...
Page 79: ...8600 Series 9 5 Thermo Scientific Section 9 Refrigeration Schematics...
Page 80: ...9 6 8600 Series Thermo Scientific Section 9 Refrigeration Schematics...
Page 81: ...8600 Series 10 1 Thermo Scientific Section 10 Electrical Schematics...
Page 82: ...10 2 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 83: ...8600 Series 10 3 Thermo Scientific Section 10 Electrical Schematics...
Page 84: ...10 4 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 85: ...8600 Series 10 5 Thermo Scientific Section 10 Electrical Schematics...
Page 86: ...10 6 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 87: ...8600 Series 10 7 Thermo Scientific Section 10 Electrical Schematics...
Page 88: ...10 8 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 89: ...8600 Series 10 9 Thermo Scientific Section 10 Electrical Schematics...
Page 90: ...10 10 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 91: ...8600 Series 10 11 Thermo Scientific Section 10 Electrical Schematics...
Page 92: ...10 12 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 93: ...8600 Series 10 13 Thermo Scientific Section 10 Electrical Schematics...
Page 94: ...10 14 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 95: ...8600 Series 10 15 Thermo Scientific Section 10 Electrical Schematics...
Page 96: ...10 16 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 97: ...8600 Series 10 17 Thermo Scientific Section 10 Electrical Schematics...
Page 98: ...10 18 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 99: ...8600 Series 10 19 Thermo Scientific Section 10 Electrical Schematics...
Page 100: ...10 20 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 101: ...8600 Series 10 21 Thermo Scientific Section 10 Electrical Schematics...
Page 102: ...10 22 8600 Series Thermo Scientific Section 10 Electrical Schematics...
Page 103: ...8600 Series 10 23 Thermo Scientific Section 10 Electrical Schematics...
Page 104: ...10 24 8600 Series Thermo Scientific Section 10 Electrical Schematics...