background image

Thermo Fisher Scientific ULT Freezers

 Operation  |   9

Operation

Initial Start Up

To start the freezer, complete the following steps:

1. Plug the freezer into the power outlet.

2. Turn the power switch ON. You can find the switch in

back of the freezer, on the bottom right.

3. Once the freezer is turned ON, the user interface will

begin a start up procedure. Once ready for operation, the
temperature is displayed on the screen.

Operation Overview

Once you have successfully completed the initial start up 
procedures, the freezer starts operating normally and the only 
actions required are:

Setting the operating and alarm set points.

Activating the CO

2

 or LN

2

 backup system if installed. For

instructions on backup settings and activating the
system, refer to 

Backup System (Optional)

.

Display

The display screen below is the default screen.

Figure 3. Display

The control panel consists of the 5 touch-point buttons 
located on the right side of the display.

Warm Alarm value - The range of the warm alarm
temperature is –40°C to within 5°C of setpoint.

Note :  

The warm alarm will be disabled for 12 hours

from a warm start condition.

Cold Alarm value - The range is –99°C to within 5°C
of setpoint.

Note :  

A setpoint change may automatically change

the warm/cold alarm setpoints as well to maintain a
minimum 5°C separation from the control setpoint.

Offset value

 - This is used for calibration. Range is

–10°C to +10°C. Default is 0.

Entering a positive offset value will yield a colder 
cabinet temperature. 

Entering a negative offset value will yield a warmer 
cabinet temperature.

Setpoint security code

 – This code is a 3-digit

numeric code. Refer to 

Setpoint Security

.

Backup system type 

(if backup system is installed)

– Set the type to either LN

2

 or CO

2

 corresponding to

the backup system that is installed.

Backup system setpoint

 (if backup system is

installed) - This setpoint indicates the temperature at
which the backup system will begin cooling the cabi-
net. It is recommended to set the backup system
setpoint at a minimum of 10°C warmer than the con-

1.

Alarm Bell

 – This icon indicates visual and 

audible alarm that accompanies various alarm 
states. Pressing the alarm bell while in an alarm 
state will snooze the audible alarm for 10 
minutes.

2.

Plus

 – Increases the value of the selected setting.

3.

Check Mark

 – Saves a change to the selected 

value.

4.

Minus

 – Decreases the value of the selected 

setting.

5.

Settings

 – The settings icon represents the 

various settings including:

Summary of Contents for Forma FDE Series

Page 1: ...Ultra Low Temperature Freezers RevcoTM RDE Series FormaTM FDE Series HERAfreezeTM HDE Series and Thermo ScientificTM TDE Series Installation and Operation 329712H02 Revision A March 2019...

Page 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Page 3: ...attery Door Opening Closing 4 Operating Standards 5 Electrical Specifications 5 Installation 6 Location 6 Leveling 6 Backup System Optional 6 Super Insulated Cabinet Construction 6 Door Operation 6 Pr...

Page 4: ...on 14 Changing Chart Paper 14 Calibration Adjustment 14 Maintenance 15 Cleaning the Condenser 15 Cleaning the Condenser Filter 15 Gasket Maintenance 15 Defrosting the Freezer 15 Battery Maintenance 15...

Page 5: ...odels 1 Models Table 1 Applicable Models Brand Model Size xxx Voltage Forma FDExxx86L 300 400 500 600 A D V Thermo Scientific TDExxx86L 300 400 500 600 A D V HERAfreeze HDExxx86L 300 400 500 600 A D V...

Page 6: ...ntamination procedures use chemically resistant gloves Use insulated gloves for handling samples and when using liquid nitrogen Before installing using or maintaining this product please be sure to re...

Page 7: ...en submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment The packaging can be stored and re used Packing List Inside the freezer cabinet is a bag...

Page 8: ...helf allow the freezer to recover to the desired set point before loading the next shelf Repeat this process until the freezer is fully loaded Battery Door Opening Closing To open the grille door pull...

Page 9: ...he nominal voltage Electrical Specifications The last character in the model number listed on the data plate identifies the electrical specifications for your unit Specific unit current rating is list...

Page 10: ...odels are equipped with an advanced assembly specifically designed for ultra low temperature freezers Features include One hand operation A front accessible lock Hasps for a standard padlock to provid...

Page 11: ...excessive frost accumulation on the inner door could eventually restrict air flow Therefore you should periodically inspect the inner door and brush away any loose frost using a stiff nylon brush Ins...

Page 12: ...273 274 STANDARDS FOR TISSUE BANKING 14th Ed Section E E3 300 E3 310 Table E3 331 Note for clarity The Bone is the Osteoarticular Graft a weight bearing allograft with intact articular surfaces consi...

Page 13: ...larm value The range is 99 C to within 5 C of setpoint Note A setpoint change may automatically change the warm cold alarm setpoints as well to maintain a minimum 5 C separation from the control setpo...

Page 14: ...will automatically return to the temperature display Warm Alarm Test Press the plus and checkmark buttons at the same time to initiate the warm alarm test During the warm alarm test the actual cabine...

Page 15: ...are showing Press the checkmark while YES is showing With a flashing YES you are required to confirm by pressing the checkmark button a second time Power down will then be complete If the checkmark b...

Page 16: ...equipped with siphon tubes for withdrawing liquid from the bottom of the cylinder CO2 cylinders must be kept at room temperature to function properly LN2 bottles are functional at any reasonable temp...

Page 17: ...ay shorten the life of the hose Start Up When the unit is started it will recognize if a backup system is installed 1 Follow the instructions in Backup System Optional to set the backup system type an...

Page 18: ...econd The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the lef...

Page 19: ...et 6 If there is freezer odor wash the interior with a solution of baking soda and warm water 7 Clean the exterior with any common household cleaner 8 Close the doors restart the freezer and reload Re...

Page 20: ...und the gasket inner doors and breakers as necessary Gasket Periodically check the gaskets around the door for punctures or tears Periodically clean the ice build up around the gasket Filter Clean the...

Page 21: ...igh relative humidity Check if the location meets requirements Maximum relative humidity 60 for temperatures within 15 C to 32 C 59 F to 90 F Excess frost build up in chamber Defrost the unit Refer De...

Page 22: ...o puncture the rubber gasket Excessive and prolonged door openings Avoid opening door for a prolonged time Gasket damage Check for punctures or tears on gasket If replacement is necessary call service...

Page 23: ...are running more often than before Ambient conditions Unit performance is directly impacted by these causes mentioned Try maintaining ambient conditions reducing load reducing door openings Once temp...

Page 24: ...01 Firmware Build Incompatible E02 Control Probe Failure E03 Heat Exchanger Probe Failure E04 Power Failure E05 Failure to Reach Setpoint E06 BUS Battery Failure E07 System Battery Failure E08 Lost Co...

Page 25: ...luded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original fi...

Page 26: ...able items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the wa...

Page 27: ...rature variation will increase Contact customer service Display will intermittently show E03 Ambient Probe Failure Wrench Yes Ambient Probe TC has malfunctioned This doesn t affect the performance of...

Page 28: ...s incompatibility that can result in modules to be non coherent Display will intermittently show E01 System Battery Failure Alarm Wrench Yes System Battery Disconnected or Failed Display will intermit...

Page 29: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Page 30: ...contact information for your area Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States Countries Sales North America 1 866 984 3766 India toll free 1800 22 8374 India 91 22 671...

Reviews: