Cleaning the Condenser

Clean the condenser at least every six months; more often if 
the laboratory area is dusty.

To clean the condenser, complete the following steps:

1. Pull the grille door open.

2. Vacuum the condenser.

3. Inspect the filter cleanliness and clean as required.

4. Close the grille door.

Cleaning the Condenser 
Filter

Clean the condenser filters every two or three months.

1. Pull the grille door open.

2. Remove the filter.

3. Shake the filter to remove loose dust, rinse the filters in

clean water, shake the excess water from the filter, and
replace the filter.

4. Close the grille door.

Gasket Maintenance

Periodically check the gaskets around the door for punctures 
or tears. Leaks are indicated by a streak of frost which forms 
at the point of gasket failure. Make sure that the cabinet is 
level (refer 

Leveling

 for leveling information).

Keep the door gaskets clean and frost free by wiping gently 
with a soft cloth.

Defrosting the Freezer

Defrost the freezer once per year or whenever the ice buildup 
exceeds 3/8”. To defrost, complete the following steps:

1. Remove all products and place in another ULT freezer.

2. Turn off the freezer.

3. Open the outer door and all inner doors.

4. Let the freezer stand with doors open for at least 24

hours. This allows both the interior and foamed
refrigeration system to warm to room temperature.

5. Dispose of the ice and wipe out any water standing in the

bottom of the cabinet.

6. If there is freezer odor, wash the interior with a solution of

baking soda and warm water.

7. Clean the exterior with any common household cleaner.

8. Close the doors, restart the freezer and reload. Refer

Initial Loading

 to follow the instructions.

Battery Maintenance

The freezer monitors the voltage status of the battery daily 
and indicates the battery’s voltage via visual and auditory 
alarm. Replace the battery as indicated by system alarms or 
as necessary per individual status evaluation. Check the 
battery connections regularly. Although not required, annual 
battery replacement is recommended to ensure proper 
battery status in the event of power failure. 

For safety, it is recommended to power off the unit and 
disconnect it from the power source before replacing the 
battery. Battery terminals are color coded in red and black. 
Ensure the corresponding colored wires are connected to the 
matching color terminals on the battery. The battery is 
installed with terminals oriented toward the condenser 
compartment or hinge side of the freezer’s outer door (refer 

Battery Specification

). With proper installation, the red wire 

should be connected to the rear battery (positive) terminal and 
the black wire to the front (common) terminal. 

Failure to properly connect the battery can damage electrical 
components and potentially hinder normal operation of the 
freezer. Consult a certified service technician if there are any 
questions or concerns about battery maintenance.

WARNING: 

Unauthorized repair of your freezer 

will invalidate your warranty. Contact Technical 
Service. See 

Contact Information

 for phone 

numbers. 

CAUTION: 

Maintenance should only be 

performed by trained personnel.

Summary of Contents for Forma FDE Series

Page 1: ...Ultra Low Temperature Freezers RevcoTM RDE Series FormaTM FDE Series HERAfreezeTM HDE Series and Thermo ScientificTM TDE Series Installation and Operation 329712H02 Revision A March 2019...

Page 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Page 3: ...attery Door Opening Closing 4 Operating Standards 5 Electrical Specifications 5 Installation 6 Location 6 Leveling 6 Backup System Optional 6 Super Insulated Cabinet Construction 6 Door Operation 6 Pr...

Page 4: ...on 14 Changing Chart Paper 14 Calibration Adjustment 14 Maintenance 15 Cleaning the Condenser 15 Cleaning the Condenser Filter 15 Gasket Maintenance 15 Defrosting the Freezer 15 Battery Maintenance 15...

Page 5: ...odels 1 Models Table 1 Applicable Models Brand Model Size xxx Voltage Forma FDExxx86L 300 400 500 600 A D V Thermo Scientific TDExxx86L 300 400 500 600 A D V HERAfreeze HDExxx86L 300 400 500 600 A D V...

Page 6: ...ntamination procedures use chemically resistant gloves Use insulated gloves for handling samples and when using liquid nitrogen Before installing using or maintaining this product please be sure to re...

Page 7: ...en submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment The packaging can be stored and re used Packing List Inside the freezer cabinet is a bag...

Page 8: ...helf allow the freezer to recover to the desired set point before loading the next shelf Repeat this process until the freezer is fully loaded Battery Door Opening Closing To open the grille door pull...

Page 9: ...he nominal voltage Electrical Specifications The last character in the model number listed on the data plate identifies the electrical specifications for your unit Specific unit current rating is list...

Page 10: ...odels are equipped with an advanced assembly specifically designed for ultra low temperature freezers Features include One hand operation A front accessible lock Hasps for a standard padlock to provid...

Page 11: ...excessive frost accumulation on the inner door could eventually restrict air flow Therefore you should periodically inspect the inner door and brush away any loose frost using a stiff nylon brush Ins...

Page 12: ...273 274 STANDARDS FOR TISSUE BANKING 14th Ed Section E E3 300 E3 310 Table E3 331 Note for clarity The Bone is the Osteoarticular Graft a weight bearing allograft with intact articular surfaces consi...

Page 13: ...larm value The range is 99 C to within 5 C of setpoint Note A setpoint change may automatically change the warm cold alarm setpoints as well to maintain a minimum 5 C separation from the control setpo...

Page 14: ...will automatically return to the temperature display Warm Alarm Test Press the plus and checkmark buttons at the same time to initiate the warm alarm test During the warm alarm test the actual cabine...

Page 15: ...are showing Press the checkmark while YES is showing With a flashing YES you are required to confirm by pressing the checkmark button a second time Power down will then be complete If the checkmark b...

Page 16: ...equipped with siphon tubes for withdrawing liquid from the bottom of the cylinder CO2 cylinders must be kept at room temperature to function properly LN2 bottles are functional at any reasonable temp...

Page 17: ...ay shorten the life of the hose Start Up When the unit is started it will recognize if a backup system is installed 1 Follow the instructions in Backup System Optional to set the backup system type an...

Page 18: ...econd The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the lef...

Page 19: ...et 6 If there is freezer odor wash the interior with a solution of baking soda and warm water 7 Clean the exterior with any common household cleaner 8 Close the doors restart the freezer and reload Re...

Page 20: ...und the gasket inner doors and breakers as necessary Gasket Periodically check the gaskets around the door for punctures or tears Periodically clean the ice build up around the gasket Filter Clean the...

Page 21: ...igh relative humidity Check if the location meets requirements Maximum relative humidity 60 for temperatures within 15 C to 32 C 59 F to 90 F Excess frost build up in chamber Defrost the unit Refer De...

Page 22: ...o puncture the rubber gasket Excessive and prolonged door openings Avoid opening door for a prolonged time Gasket damage Check for punctures or tears on gasket If replacement is necessary call service...

Page 23: ...are running more often than before Ambient conditions Unit performance is directly impacted by these causes mentioned Try maintaining ambient conditions reducing load reducing door openings Once temp...

Page 24: ...01 Firmware Build Incompatible E02 Control Probe Failure E03 Heat Exchanger Probe Failure E04 Power Failure E05 Failure to Reach Setpoint E06 BUS Battery Failure E07 System Battery Failure E08 Lost Co...

Page 25: ...luded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original fi...

Page 26: ...able items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the wa...

Page 27: ...rature variation will increase Contact customer service Display will intermittently show E03 Ambient Probe Failure Wrench Yes Ambient Probe TC has malfunctioned This doesn t affect the performance of...

Page 28: ...s incompatibility that can result in modules to be non coherent Display will intermittently show E01 System Battery Failure Alarm Wrench Yes System Battery Disconnected or Failed Display will intermit...

Page 29: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Page 30: ...contact information for your area Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States Countries Sales North America 1 866 984 3766 India toll free 1800 22 8374 India 91 22 671...

Reviews: