2  |   Safety Considerations

Thermo Fisher Scientific ULT Freezers

Safety Considerations

In this manual, the following symbols and conventions are 
used:

Below are important safety precautions that apply to this 
product:

This symbol used alone indicates important 
operating instructions which reduce the risk of 
injury or poor performance of the unit.

CAUTION: 

This symbol, in the context of a 

CAUTION, indicates a potentially hazardous 
situation which if not avoided could result in 
minor to moderate injury or damage to the 
equipment.

WARNING: 

This symbol indicates potentially 

hazardous situations which, if not avoided, could 
result in serious injury or death.

WARNING: 

This symbol indicates situations 

where dangerous voltages exist and potential for 
electrical shock is present.

The snowflake symbol indicates extreme low 
temperatures and high risk of frostbite. Do not 
touch bare metal or samples with unprotected 
body parts.

This symbol indicates a need to use gloves during 
the indicated procedures. If performing 
decontamination procedures, use chemically 
resistant gloves. Use insulated gloves for handling 
samples and when using liquid nitrogen.

Before installing, using or maintaining this 
product, please be sure to read this manual and 
product warning labels carefully. Failure to follow 
these instructions may cause this product to 
malfunction, which could result in injury or 
damage.

Use this product only in the way described in the 
product literature and in this manual. Before using 
it, verify that this product is suitable for its 
intended use. If this equipment is used in a 
manner not specified by the manufacturer, the 
protection provided by the equipment may be 
impaired.

Do not modify system components, especially 
the controller. Use OEM exact replacement 
equipment or parts. Before use, confirm that the 
product has not been altered in any way.

WARNING: 

Your unit must be properly grounded 

in conformity with national and local electrical 
codes. Never connect the unit to overloaded 
power sources.

WARNING: 

Disconnect the unit from all power 

sources before cleaning, troubleshooting, or 
performing other maintenance on the product or 
its controls.

WARNING: 

“Caution, risk of fire”. This unit is 

charged with hydrocarbon refrigerants.

Summary of Contents for Forma FDE Series

Page 1: ...Ultra Low Temperature Freezers RevcoTM RDE Series FormaTM FDE Series HERAfreezeTM HDE Series and Thermo ScientificTM TDE Series Installation and Operation 329712H02 Revision A March 2019...

Page 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Page 3: ...attery Door Opening Closing 4 Operating Standards 5 Electrical Specifications 5 Installation 6 Location 6 Leveling 6 Backup System Optional 6 Super Insulated Cabinet Construction 6 Door Operation 6 Pr...

Page 4: ...on 14 Changing Chart Paper 14 Calibration Adjustment 14 Maintenance 15 Cleaning the Condenser 15 Cleaning the Condenser Filter 15 Gasket Maintenance 15 Defrosting the Freezer 15 Battery Maintenance 15...

Page 5: ...odels 1 Models Table 1 Applicable Models Brand Model Size xxx Voltage Forma FDExxx86L 300 400 500 600 A D V Thermo Scientific TDExxx86L 300 400 500 600 A D V HERAfreeze HDExxx86L 300 400 500 600 A D V...

Page 6: ...ntamination procedures use chemically resistant gloves Use insulated gloves for handling samples and when using liquid nitrogen Before installing using or maintaining this product please be sure to re...

Page 7: ...en submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment The packaging can be stored and re used Packing List Inside the freezer cabinet is a bag...

Page 8: ...helf allow the freezer to recover to the desired set point before loading the next shelf Repeat this process until the freezer is fully loaded Battery Door Opening Closing To open the grille door pull...

Page 9: ...he nominal voltage Electrical Specifications The last character in the model number listed on the data plate identifies the electrical specifications for your unit Specific unit current rating is list...

Page 10: ...odels are equipped with an advanced assembly specifically designed for ultra low temperature freezers Features include One hand operation A front accessible lock Hasps for a standard padlock to provid...

Page 11: ...excessive frost accumulation on the inner door could eventually restrict air flow Therefore you should periodically inspect the inner door and brush away any loose frost using a stiff nylon brush Ins...

Page 12: ...273 274 STANDARDS FOR TISSUE BANKING 14th Ed Section E E3 300 E3 310 Table E3 331 Note for clarity The Bone is the Osteoarticular Graft a weight bearing allograft with intact articular surfaces consi...

Page 13: ...larm value The range is 99 C to within 5 C of setpoint Note A setpoint change may automatically change the warm cold alarm setpoints as well to maintain a minimum 5 C separation from the control setpo...

Page 14: ...will automatically return to the temperature display Warm Alarm Test Press the plus and checkmark buttons at the same time to initiate the warm alarm test During the warm alarm test the actual cabine...

Page 15: ...are showing Press the checkmark while YES is showing With a flashing YES you are required to confirm by pressing the checkmark button a second time Power down will then be complete If the checkmark b...

Page 16: ...equipped with siphon tubes for withdrawing liquid from the bottom of the cylinder CO2 cylinders must be kept at room temperature to function properly LN2 bottles are functional at any reasonable temp...

Page 17: ...ay shorten the life of the hose Start Up When the unit is started it will recognize if a backup system is installed 1 Follow the instructions in Backup System Optional to set the backup system type an...

Page 18: ...econd The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the lef...

Page 19: ...et 6 If there is freezer odor wash the interior with a solution of baking soda and warm water 7 Clean the exterior with any common household cleaner 8 Close the doors restart the freezer and reload Re...

Page 20: ...und the gasket inner doors and breakers as necessary Gasket Periodically check the gaskets around the door for punctures or tears Periodically clean the ice build up around the gasket Filter Clean the...

Page 21: ...igh relative humidity Check if the location meets requirements Maximum relative humidity 60 for temperatures within 15 C to 32 C 59 F to 90 F Excess frost build up in chamber Defrost the unit Refer De...

Page 22: ...o puncture the rubber gasket Excessive and prolonged door openings Avoid opening door for a prolonged time Gasket damage Check for punctures or tears on gasket If replacement is necessary call service...

Page 23: ...are running more often than before Ambient conditions Unit performance is directly impacted by these causes mentioned Try maintaining ambient conditions reducing load reducing door openings Once temp...

Page 24: ...01 Firmware Build Incompatible E02 Control Probe Failure E03 Heat Exchanger Probe Failure E04 Power Failure E05 Failure to Reach Setpoint E06 BUS Battery Failure E07 System Battery Failure E08 Lost Co...

Page 25: ...luded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original fi...

Page 26: ...able items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the wa...

Page 27: ...rature variation will increase Contact customer service Display will intermittently show E03 Ambient Probe Failure Wrench Yes Ambient Probe TC has malfunctioned This doesn t affect the performance of...

Page 28: ...s incompatibility that can result in modules to be non coherent Display will intermittently show E01 System Battery Failure Alarm Wrench Yes System Battery Disconnected or Failed Display will intermit...

Page 29: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...