3.
Pull the filter cartridge from the right rear underside of the scanner.
4.
Remove the old filter from the cartridge, then replace with a new filter.
5.
Return the filter cartridge to the right rear underside of the scanner, then
reconnect the power cord.
6.
Power on the workstation when ready to resume use of the scanner.
Lambda LS xenon arc lamp
If the lamp does not power on when the power is turned on, check the line power fuse
to see if it has blown. The fuse is in a pry-out holder on the power input module on
the back of the lamp. A spare fuse is also provided in the power input module.
WARNING!
Replace fuse with the same type and rating: 5 Amp, 250 V, 5 x
20 mm, time delay fuse (EIC 60127-2).
1.
Unplug power cord from the back of the Lambda LS.
The slot is visible just under the edge of the fuse holder.
2.
Use a small screwdriver, then pry the holder straight out of the power entry
module.
The fuse that is readily visible in the fuse holder when you take it out is the one
that is "active" when the holder is installed. A spare fuse is also stored within the
fuse holder, as shown in the following diagram.
Replace the
Lambda LS fuse
Chapter 3
Maintain the Instrument
Lambda LS xenon arc lamp
3
18
GeneTitan
™
Scanner Use and Maintenance Guide