background image

Thermo Scientific Jewett CTF306 Blood Plasma Freezers

Installation and Operation

TROUBLESHOOTING SERVICE CHART

 

 

SYMPTOM 

 

POSSIBLE CAUSE

 

POSSIBLE CORRECTIVE STEP

 

1. Excessive load.  

1. Reduce load or add additional equipment. 

High suction 
pressure.

 

2. Expansion valve overfeeding. 

2. Check remote bulb.  Adjust superheat. 

1. Insufficient refrigerant in system.  

1. Check for leaks.  Repair and add charge. 

2. Restriction in refrigeration system.  Most 
notably the liquid line filter drier or 
capillary.  

2. Determine location and remove. 

Low suction 

pressure. 

3. Expansion valve malfunctioning. 

3. Check and reset for proper superheat. 

1. Expansion valve passing excess 
refrigerant or is oversized.  

1. Readjust valve or replace with smaller 
valve. 

2. Expansion valve stuck open. 

2. Clean valve of foreign particles, and 
replace if necessary. 

3. Evaporator fan motors not running. 

3. Determine reason and correct.  Check 
door switch.  

Suction line 
frosted or 
sweating. 

4. System overcharged with refrigerant. 

4. Correct the charge. 

1. Restriction in liquid line filter drier. 

1. Determine location and remove. 

Liquid line 

frosted or 
sweating 

2. Liquid line shutoff valve partially closed.  2. Open valve. 

1. Defrost duration too long. 

1.Adjust defrost termination. 

2. Fan delay not delaying fans after defrost 

period. 

2. Defective fan delay thermostat.  Replace. 

3. Defective timer. 

3. Replace. 

Ice 
accumulating 

on ceiling 
around 
evaporator 
and/or on fan 

guards or 
blades. 

4. Too many defrost cycles per day. 

4. Adjust timer for less defrost cycles. 

1. Coil temperature not getting above 
freezing point during defrost. 

1. Check heater operation, or hot gas 
solenoid valve. 

2. Not enough defrost cycles per day. 

2. Adjust timer for more defrost cycles. 

3. Defrost cycle too short. 

3. Adjust timer for longer defrost cycle. 

4. Poor door seal. 

4. Adjust door latch, install new gasket. 

Evaporator 
coil not 

clearing of 
frost during 
defrost cycle. 

5. Defective timer or defrost thermostat. 

5. Replace defective component. 

1. Defective heater. 

1. Replace heater. 

2.Unit not pitched properly. 

2. Check and adjust if necessary. 

3. Drain line plugged. 

3. Clean drain line. 

4. Defective drain line heater. 

4. Replace heater. 

5. Poor contact between drain pan and 

heater element. 

5. Repair. 

Ice 
accumulating 
in drain pan. 

6. Defective timer or defrost thermostat. 

6. Replace defective component. 

NOTE:

 

 Jewett Refrigerators and Freezers are designed to operate in areas that are heated to 60

°

 F (15.6

°

  C).  Installation in 

unheated areas may require a low temperature compressor protection kit for satisfactory operation. 

 

 
12

Summary of Contents for Jewett CTF306

Page 1: ...Installation and Operation Manual Thermo Scientific Jewett CTF306 Blood Plasma Freezers...

Page 2: ...2009 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries...

Page 3: ...peration Do not adjust the refrigerant valves or the temperature control 7 Use of electrical characteristics other than those specified on the serial plate will cause permanent damage to the mechanism...

Page 4: ...2...

Page 5: ...3...

Page 6: ...ormed according to the temperature measured by the thermostat probe with a positive differential from the set point if the temperature increases and reaches set point plus differential the compressor...

Page 7: ...2 100 P2 CCt Compressor ON time during continuous cycle 0 0 24 0h res 10min Allows to set the length of the continuous cycle compressor stays on without interruption for the CCt time Can be used for i...

Page 8: ...The alarm stops just after the digital input is de activated 8 3 SERIOUS ALARM MODE i1F bAL When the digital input is activated the unit will wait for did delay before signalling the CA alarm message...

Page 9: ...P Third probe presence n not present Y pres n Pr2 O3 Third probe calibration 12 12 C 120 120 F 0 0 0 Pr2 P4P Fourth probe presence n not present Y pres n Pr2 O4 Fourth probe calibration 12 12 C 120 12...

Page 10: ...ce to removal of the dollar bill test strip should be felt Check the perimeter of the door If the latch is loose see latch adjusting information 5 Replace cover plate A Note To accomplish different of...

Page 11: ...line solenoid will not open 10 Repair or replace coil 1 Low voltage to unit 1 Determine reason and correct 2 Start capacitor failure or wrong 2 Replace start capacitor 3 Run capacitor failure or wrong...

Page 12: ...nt 1 Leak check and repair 2 Control contacts stuck or frozen 2 Clean contacts or replace control 3 Refrigerated air space has an excessive load 3 Determine reason and correct 4 Dirty Condenser 4 Clea...

Page 13: ...lve 1 Defrost duration too long 1 Adjust defrost termination 2 Fan delay not delaying fans after defrost period 2 Defective fan delay thermostat Replace 3 Defective timer 3 Replace N Ice accumulating...

Page 14: ...unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elet...

Page 15: ...e information can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase...

Page 16: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 306650H01 Rev D...

Reviews: