background image

Thermo Scientific Jewett CTF306 Blood Plasma Freezers

Installation and Operation

1

Introduction

This product is a complete packaged unit ready to operate when 
plugged into an electrical source.  It is not necessary to have a 
refrigeration serviceman set the unit in operation.  The unit has 
been tested prior to shipment.  Read all the instructions before 
proceeding with installation.. 

2

Inspection for Damage

Uncrate the product and remove all packing and crating 
materials.  Inspect the product and parts thoroughly for damage 
or missing parts.  File all claims for damage with the 
transportation company immediately.  Do not file claims with the 
manufacturer.

3

Installation Instruction

1. The cabinet will pass through a standard 30" door opening.

2. Inspect the interior, exterior and mechanical equipment for 

special instruction tags fastened at various points.

3. Move the freezer into the desired location.  Make sure the 

bottom of the refrigerator is evenly supported.  Thin shims 
under the points of rest can be used to equalize the 
distribution of weight.  If the cabinet sets on an uneven floor, 

a slight rocking or vibration might result when the 
condensing unit is set in operation.

4. Make certain the cabinet is located so the grill panel is 

unobstructed.

5. An automatic condensate evaporator eliminates the necessity 

for a floor drain.  For proper elimination of condensation, 
wafers should be placed in the condensate tray located in the 
rear of the mechanical compartment in accordance with the 
instructions packed with the wafers.

6. The condensing unit is shipped with all service valves open 

and ready for operation.  Do not adjust the refrigerant valves 
or the temperature control. 

7. Use of electrical characteristics, other than those specified on 

the serial plate will cause permanent damage to the 
mechanism. The serial plate is located inside the freezer at 
the top of the right hand wall.

8. Operate the freezer for several hours to allow the unit to 

reach normal operating temperatures before storing product. 

NOTE:

 

Erratic operation on initial start-up does not indicate a 

faulty control.  When normal operating temperature is reached, 

the condensing unit will cycle regularly.

4

Maintenance Instructions

1. Frequent cleaning of the interior and exterior with water and 

a good fungicidal detergent that eliminates harmful bacteria, 
stains and other foreign matter will keep the unit fresh and 
new looking.

2. Shelves and/or drawers should be removed from the cabinet 

and thoroughly scrubbed. Clean door gasket periodically.

3. The unit cooler fan operates continuously when the door is 

closed and requires no lubrication. 

4. The condensing unit needs no oil or other lubrication.  The 

finned condenser can become clogged with lint or dust.  The 
openings between the fins should be kept clean.  A vacuum 
cleaner or small test tube brush works well for this purpose.  

This should be accomplished on an annual basis.  Failure to 
keep the condenser fins free of dirt and lint will result in 
erratic operation and may damage the refrigeration system.

5. Annual inspection of the mechanical refrigeration equipment 

by a qualified serviceman is recommended.  A qualified 
mechanic can frequently make adjustments that will prevent 
future breakdown.

6. If the system is opened for any reason, extreme care should 

be taken to prevent the entry of moisture-bearing air.  A new 
drier should be installed in the lines when the system is 
closed.

 
1

Summary of Contents for Jewett CTF306

Page 1: ...Installation and Operation Manual Thermo Scientific Jewett CTF306 Blood Plasma Freezers...

Page 2: ...2009 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries...

Page 3: ...peration Do not adjust the refrigerant valves or the temperature control 7 Use of electrical characteristics other than those specified on the serial plate will cause permanent damage to the mechanism...

Page 4: ...2...

Page 5: ...3...

Page 6: ...ormed according to the temperature measured by the thermostat probe with a positive differential from the set point if the temperature increases and reaches set point plus differential the compressor...

Page 7: ...2 100 P2 CCt Compressor ON time during continuous cycle 0 0 24 0h res 10min Allows to set the length of the continuous cycle compressor stays on without interruption for the CCt time Can be used for i...

Page 8: ...The alarm stops just after the digital input is de activated 8 3 SERIOUS ALARM MODE i1F bAL When the digital input is activated the unit will wait for did delay before signalling the CA alarm message...

Page 9: ...P Third probe presence n not present Y pres n Pr2 O3 Third probe calibration 12 12 C 120 120 F 0 0 0 Pr2 P4P Fourth probe presence n not present Y pres n Pr2 O4 Fourth probe calibration 12 12 C 120 12...

Page 10: ...ce to removal of the dollar bill test strip should be felt Check the perimeter of the door If the latch is loose see latch adjusting information 5 Replace cover plate A Note To accomplish different of...

Page 11: ...line solenoid will not open 10 Repair or replace coil 1 Low voltage to unit 1 Determine reason and correct 2 Start capacitor failure or wrong 2 Replace start capacitor 3 Run capacitor failure or wrong...

Page 12: ...nt 1 Leak check and repair 2 Control contacts stuck or frozen 2 Clean contacts or replace control 3 Refrigerated air space has an excessive load 3 Determine reason and correct 4 Dirty Condenser 4 Clea...

Page 13: ...lve 1 Defrost duration too long 1 Adjust defrost termination 2 Fan delay not delaying fans after defrost period 2 Defective fan delay thermostat Replace 3 Defective timer 3 Replace N Ice accumulating...

Page 14: ...unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elet...

Page 15: ...e information can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase...

Page 16: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 306650H01 Rev D...

Reviews: