Operation
Initial Startup
Thermo Scientific
Jewett Pass-Thru Blood Bank Refrigerators
13
Chapter 5
Operation
Initial Startup
•
Make sure that the unit is free of all wood or cardboard shipping
materials, both inside and outside.
•
Plug unit into a power outlet on a dedicated circuit.
•
Turn unit off using power switch before unplugging or servicing.
•
Allow unit to reach operating temperature before storing contents.
Performance
The refrigerator is factory preset to operate within a +2 °C to +4 °C
temperature range.
The refrigerator temperature will be shown on the digital display at the top
of the unit.
Defrost
The defrosting process initiates automatically in response to a built-in timer.
All models are set for one defrost cycle every 1 hour. The maximum length
of the defrost cycle is 15 minutes. The cycle terminates automatically if
during defrost the evaporator coil temperature e4°C.
Note
•
All models have been provided with an electric condensate evaporator
located at the cabinet bottom. Periodic cleaning of the condensate pan
will ensure proper operation. Refer to image below for location of
condensate pan.
Summary of Contents for Jewett Pass-Thru BBR25D-1B
Page 4: ...Table of Contents 2 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 8: ...Inspection for Damage 6 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 14: ...Installation Wiring 12 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 32: ...Troubleshooting 32 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 34: ...Accessories 34 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 36: ...36 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...