Hinge and Latch Adjustment
Hinge Adjustment
Thermo Scientific
Jewett Pass-Thru Blood Bank Refrigerators
19
Chapter 7
Hinge and Latch
Adjustment
Hinge Adjustment
1. Remove exterior hinge cover plate by opening the refrigerator door.
Place a screwdriver under interior portion of cover and gently pry
cover outward. Pull straight out. Close cover.
2. Loosen the three (3) screws "B" which hold adjusting plate "C" in
position.
3. To tighten gasket seal, place hand against exterior of door near
hinges; gently press in on door so gasket sits firmly against cabinet
face. Tighten screws "B".
4. When adjustment is complete, it hinges are adjusted so gasket seal is
too tight, door will tend to spring open. Door must be readjusted.
To test gasket seal, insert a dollar bill (or piece of paper of similar
size) between the gasket and the cabinet opening. Close door; a
slight resistance to removal of the dollar bill (test strip) should be
felt. Check the perimeter of the door. If the latch is loose, see
latch-adjusting information.
5. Replace cover plate "A."
Note
To accomplish different offsets, shims are utilized. If replacing hinges,
make sure to reuse any shims furnished on the cabinet.
Summary of Contents for Jewett Pass-Thru BBR25D-1B
Page 4: ...Table of Contents 2 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 8: ...Inspection for Damage 6 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 14: ...Installation Wiring 12 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 32: ...Troubleshooting 32 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 34: ...Accessories 34 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...
Page 36: ...36 Jewett Pass Thru Blood Bank Refrigerators Thermo Scientific...