Installation
Safety Instructions
Thermo Scientific
Jewett Pass-Thru Blood Bank Refrigerators
9
Chapter 4
Installation
Safety Instructions
•
Only plug into a grounded outlet
•
Do not remove ground prong
•
Do not use a 3 prong to 2 prong adapter
•
Do not use an extension cord
•
Disconnect power before servicing
•
Do not operate with any panels or parts removed
Location
1. Measure all doorways and passageways for clearance before moving
cabinet. These units will pass through a standard 30" x 84” door
opening.
2. Inspect the interior, exterior and mechanical equipment for special
instruction tags fastened at various points.
3. Move the unit into the desired location. Make sure the bottom is
evenly supported and the floor is level. Thin shims under the points
of rest can be used to equalize the distribution of weight. If the
cabinet sets on an uneven floor, a slight rocking or vibration might
result when the condensing unit is set in operation.